KEI 7059 Instruction 电路图.pdf
《KEI 7059 Instruction 电路图.pdf》由会员分享,可在线阅读,更多相关《KEI 7059 Instruction 电路图.pdf(19页珍藏版)》请在收音机爱好者资料库上搜索。
1、Model 7059 Low Voltage Scanner Plug-h Card Instruction Manual Publication Date: February 1989 Document Number: 7059-901-01 Rev. E RadioFans.CN 收音机爱 好者资料库 WARRANTY RadioFans.CN 收音机爱 好者资料库 SPECIFICAITONS MODEL 7059 LOW VOLTAGE SCANNER CARD CHANNELS PER CARD: 10. CONTACT CONFIGURATION: Z-pole Form A, c
2、ommon guard connection. CONNECTOR TYPE: Screw terminal, No. 18 AWG maximum wire size. RELAY DRIVE CURRENT: UmA per relay typical. MAXIMUM SIGNAL LEVEL: ZOOV, lOOmA, 2VA peak (resistive load). CONTACT LIFE: IO 8 10 closuvs (at maximum signal level). CONTACT RESISTANCE: Zfl to rated life. CONTACT POTE
3、NTIAL: l 6 Common Mode Isolation Test Setup . 9 9 Model 7058 Component Location Drawing . 13 10 Model 7059 Schematic Diagram . 15 LIST OF TABLES 1 Recommended Test Equipment, . 5 2 Model 7059 Replaceable Parts List. . 12 SECTION 1 GENERAL INFORMATION 1.1 INTRODUCTION The Model 7059 is a low-voltage
4、scanner plug-in card which is field- installable in the Model 705 and Model 706 scanner mainframes. The Model 7059 will switch up to 10 channels. For low-level transducer and thermo- couple output switching the Model 7059 is designed to introduce a minimum of thermal voltage error I l less than 2fI
5、at end of life. C. A guard surrounds all analog signal paths. 2. Installation- Refer to the Model 705 and Model 706 Instruction Manuals for scanner card installation instructions. 2.3 OPERATING CONSIDERATIONS 1. Signal Level- 1OV peak, 1OmA peak with a resistive load for expected life. Absolute maxi
6、mum peak instantaneous rating: 2OOV. lOOmA, or 2VA with a resistive load. 2. Isolation-Guarded interchannel resistance is nominally greater than 10t2fI and less than 1OpF at room temperature, at less than 70% relative humidity. 3. Maximum Levels-200V peak between signal line pairs or from signal lin
7、es to guard or mainframe (digital) common. 4. Operating Environment-O0 to 50C up to 35OC at 70% relative humidity. Figure 1. Plug-In Card Installation and Assembly 3 2.4 OPERATING HINTS The clamp-type screw terminals will accept up to #19 AWG wire. In order to ensure thermal offset less than l$/ sol
8、id copper wires should be used throughout the measurement hook-up. To most effectively eliminate error voltages produced by leakage currents. the GUARD terminal should be connected to the reference connection on the channel which is most sensitive to error (that is, the one which has the lowest sour
9、ce voltage versus the highest series resistance). For instance, two sources are connected to the Model 7059; one having an equivalent source of 1V with a series resistance of lOOk0, and the other having a source voltage of .lV with a lkn series resistance. The lOOk source resistance will develop 100
10、 times the error voltage as the lkt3 source resistance, Its source voltage however, is only ten times as large on the lV/lOOkn channel. The GUARD would therefore be connected to the reference point of this channel lsee Figure 21. Reactances in the system will cause transients during switching. These
11、 should not exceed the ratings given in the specifications. If they do, relay contact life will be degraded. Figure 2. How To Use GUARD Connection 4 SECTION 3 MAINTENANCE 3.1 INTRODUCTION This section contains a performance verification procedure. Since there is no calibration adjustments, no recali
12、bration is necessary. Recommended main- tenance would include inspection of the scanner plug-in board and card- edge connectors to ensure good electrical contact. In industrial environ- ments annual cleaning using dry nitrogen gas and FreonO TMS or TE is recommended. The verification procedure shoul
13、d be performed every 12 months or at the time maintenance is performed on the mainframe. 3.2 REQUIRED TEST EQUIPMENT Recommended test equipment is given in Table 1, Test equipment other than recommended may be substituted if specifications equal or exceed the stated characteristics. Table 1. Recomme
14、nded Test Equipment Thermal Cable I i Hinimum Specification !ero drift less than 150nV. TTL compatible square wave to 10Hz. and 1 pulse per 30 min. ItI Sensitivity rriggered Sweep TEK IO.OOV, 1OOVDC KI 1UOA Sensitivity KI dFG i Model KI 181 KI 1488 KI 1506 H-P 70358 705 195 1641 000 seriu 230 i14 or
15、 61 3.3 PERFORMANCE VERIFICATION This section gives procedures needed to verify that the operation of the Model 7059 is within specifications. The thermal offset should be carried out in a temperature controlled environment of 23O f 1C. Other tests may be performed within environmental limits given
16、in the specifications. 3.3.1 Thermal Offset Test 1. The input of the Model 181 should be shorted with the Model 1488 short- ing plug. After power is applied, the Model 181 should be allowed to stabilize for four hours. During this time the analog output of the Model 5 181 should be connected to the
17、chart recorder end a recording made to establish a baseline for the voltage measurements lsee Figure 3). 2. The 10 input connections on the Model 7059 plug-in card should be shorted together between HI end LO with a short piece of solid copper wire. That is, connect all the HI and LO terminals toget
18、her es shown in Figure 4. (The OUTPUT terminals of the Mods1 7059 should be connected to short copper wires to facilitate connecting the alligator clips on the Model 1506 cable.) With the Model 1506 cable attached to the output, the plug-in card should then be inserted into the mainframe and power a
19、p- plied to the mainframe. Set the Model 705 to the reset mods and allow it to warm up along with the Model 181. 3. After the warm up time has elapsed and a stable baseline is visible on the chart recorder, the shorting plug on the Model 181 should be removed and the Model 1506 cable from the Model
20、7059 connected in its place. Set the Model 705 scanner mainframe to Channel 1, Channel mode and the Single program mode. The scanner mainframe should now be clocked et a rate of 1 channel every 15 (900 second interval time) minutes. The thermal EMF generated in each channel will be visible as deflec
- 配套讲稿: