1、1164.4556.82-01- 1 Test and Measurement Division Service Manual Instrument SPECTRUM ANALYZER R -5 dBm; -10 dBm; -15 dBm; -20 dBm; -25 dBm Measurement:Set marker to peak: - MKR ? : PEAK Checking Frequency Response Test equipment:- N cable (refer to Measurement Equipment, item 14) Frequencyup to 3 GHz
2、 Maximum attenuation 0.2 dB Test setup:Connect tracking generator output to RF input of the FSP. FSP settings:- PRESET - NETWORK - BW : RES BW MANUAL : 1 kHz - AMPT : RF ATTEN MANUAL : 10 dB - AMPT : 0 dBm - BW : RES BW MANUAL : 1 kHz - SWT : MANUAL : 2 s Range 1: - FREQ : START : 9 kHz - FREQ : STO
3、P : 100 kHz Range 2: - FREQ : START : 100 kHz - FREQ : STOP : 1 MHz Range 3: - FREQ : START : 1 MHz - FREQ : STOP : 2 GHz Range 4: - FREQ : START : 2 GHz - FREQ : STOP : 3 GHz - NETWORK : SOURCE POWER level Values for level : 0 dBm; -10 dBm; -20 dBm Performance TestFSP 1166.4556.821.22E-1 Measuremen
4、t:Set marker to peak: - MKR ? : PEAK Set marker to minimum value: - MKR ? : MIN PEAK Checking Modulation Checking I/Q Modulation Test equipment:- ADS (refer to Measurement Equipment, item 16) - Spectrum Analyzer (refer to Measurement Equipment, item 15) - 2 Voltmeters (refer to Measurement Equipment
5、, item 17) - 2 BNC T pieces Test setup:? Apply the I and Q outputs of ADS to the I and Q inputs of the analyzer. Check the ADS voltages in parallel via T pieces. ? Connect spectrum analyzer to the TG output. ADS settings:? Set DC voltage (offset) to 0 mV 3 mV AC voltage: V(pp) = 1 V 10 mV Frequency:
6、 200 kHz Phase (I/Q) = 90 degrees Spectrum analyzer settings:- FREQUENCY : 1 GHz - SPAN : 1 MHz - REF : REF LEVEL : 0 dBm FSP settings:- FREQUENCY : 1 GHz - SPAN : 0 MHz - REF : REF LEVEL : 0 dBm - NETWORK : SOURCE POWER 0 dBm - MODULATION EXT I/Q Measurement of residual carrier: ? Switch off I and
7、Q channels on ADS. Voltmeter display: DC voltage 3 mV AC voltage 12 dB Test setup:? Connect RF outputs of the signal generators via 10 dB attenuators to the inputs of the 3 dB coupler. ? Connect output of the 3 dB coupler to RF input of the FSP. Signal generator settings (both generators): - Frequen
8、cy: generator 1fg1 = fin 50 kHz generator 2fg2 = fin + 50 kHz Refer to Performance Test Report table for values of fin. ? Adjust the output level of signal generators for an input level at the FSP of -20 dBm. FSP settings:- PRESET - AMPT : RF ATTEN MANUAL : 0 dB - AMPT : NEXT : ELEC ATTEN MANUAL : 0
9、 dB - AMPT : -10 dBm - SPAN : 500 kHz - BW : RES BW MANUAL : 3 kHz - FREQ : CENTER : fin Refer to Performance Test Report table for values of fin. Measurement:? MKR FCTN : TOI Evaluation:The third order intercept point (T.O.I) referenced to the input signal is displayed in the marker field by the re
10、ading TOI. FSPPerformance Test 1166.4556.821.33E-1 Checking RF Attenuator (with Option B25) Test equipment:- Signal generator (refer to Measurement Equipment, item 3) Frequency128 MHz Maximum level 10 dBm - Step attenuator (refer to Measurement Equipment, item 11) Frequency128 MHz Attenuation0 to 80
11、 dB in 5 dB steps Attenuation accuracy 0.1 dB Test setup: ? Connect RF output of the signal generator to RF input of the step attenuator. ? Connect RF output of the step attenuator to RF input of the FSP. Signal generator settings:- Frequency128 MHz - Level10 dBm Step attenuator settings:Attenuation
12、70 dB FSP settings:- PRESET - FREQ : CENTER : 128 MHz - SPAN : 500 Hz - BW : RES BW MANUAL : 1 kHz - TRACE : DETECTOR : RMS - BW : VIDEO BW MANUAL : 100 Hz - AMPT : RF ATTEN MANUAL : 10 dB - AMPT : -35 dBm Reference measurement: ? Set marker to peak of signal: - MKR ? : PEAK ? Set reference to peak
13、of signal: - MKR : REFERENCE FIXED Measurement Step attenuator settings:Attenuation 80 dB - aFSP Refer to table below for values of aATT. FSP settings:- AMPT : RF ATTEN MANUAL : aFSP - AMPT : -45 dBm + aFSP dBm - MKR ? : PEAK Refer to table below for values of a FSP , a ATT and reference level. Eval
14、uation:The difference between the level of the input signal of the FSP and the reference (at 10 dB RF-Attenuation) is displayed in the marker field by the reading Delta T1 FXD. aATT in dB8075706560555045403530252015105 aFSP in dB051015202530354045505560657075 Ref. level in dBm -45-40-35-30-25-20-15-
15、10-5051015202530 Performance TestFSP 1166.4556.821.34E-1 Checking Electronic Attenuator Accuracy Test equipment:- Signal generator (refer to Measurement Equipment, item 3) Frequency128 MHz Maximum level 0 dBm - Step attenuator (refer to Measurement Equipment, item 11) Frequency128 MHz Attenuation0 t
16、o 40 dB in 5 dB steps Attenuation accuracy 0.1 dB Test setup:? Connect RF output of the signal generator to RF input of the step attenuator. ? Connect RF output of the step attenuator to RF input of the FSP. Signal generator settings:- Frequency128 MHz - Level0 dBm Step attenuator settings:Attenuati
17、on40 dB FSP settings:- PRESET - FREQ : CENTER : 128 MHz - SPAN : 500 Hz - BW : RES BW MANUAL : 1 kHz - TRACE : DETECTOR : RMS - BW : VIDEO BW MANUAL : 100 Hz - AMPT : RF ATTEN MANUAL : 10 dB - AMPT : - 30 dBm Reference measurement:? Set marker to peak of signal: - MKR ? : PEAK ? Set reference to pea
18、k of signal: - MKR : REFERENCE FIXED Measurement Step attenuator settings:Attenuation 40 dB - aFSP Refer to table below for values of aATT. FSP settings:- AMPT : NEXT : ELEC ATTEN MANUAL : aFSP - AMPT : -30 dBm + aFSP dBm - MKR ? : PEAK Refer to table below for values of a FSP , a ATT and reference
19、level. Evaluation:The difference between the level of the input signal of the FSP and the reference (at 10 dB RF-Attenuation) is displayed in the marker field by the reading Delta T1 FXD. aATT40 dB35 dB20 dB25 dB20 dB15 dB10 dB aFSP0 dB5 dB10 dB15 dB20 dB25 dB30 dB Reference level-30 dBm-25 dBm-20 d
20、Bm-15 dBm-10 dBm-5 dBm0 dBm FSPPerformance Test Report 1166.4556.821.35E-1 Performance Test Report FSP Table 1-1Performance Test Report ROHDE FSP7.3.42 Replacing 7/8 GHz YIG Unit A161; FSP7 .3.43 Replacing Diplexer (30 GHz) A230; FSP13/30 .3.47 Contents - RepairFSP 1164.4556.82I-3.2E-1 Replacing 13/