1、MODEL 602 INSTRUCTION MANUAL Instruction Manual Model 602 Solid-State Electrometer Publication Date: June 1963 Document Number: 29111 Rev. C Contains Operating and Servicing Information for the Model 602 Solid-State Electrometer 01975, Keithley Instruments, Inc. Cleveland, Ohio, U.S.A. Document Numb
2、er 29111 RadioFans.CN 收音机爱 好者资料库 WARRANTY Keithley Instruments. Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment. Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
3、rechargeable batteries, diskettes, and documentation. During the warranty period, we will, at our option, either repair or replace any product that proves to be defective. To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio
4、. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products we warranted for the balance of the origi- nal warranty per
5、iod, or at least 90 days. LIMITATION OF WARRANTY This warranty does not apply to defects resulting from pmduct modification without Keithleys express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from batte
6、ry leakage, or problems arising from normal wear or failure to follow instructions. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYERS SOLE AND EXCLUSIVE RE
7、MEDIES. NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDI- RECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SU
8、CH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY. Keithley Instruments, Inc. - 28775 Aurora Road * Cleveland, OH 44139 * 216-248-0400 *Fax: 216-248-6168 * RadioFan
9、s.CN 收音机爱 好者资料库 SPECIFICATIONS AS A VOLTMETER: RANGE: .OOlV full scale to 1OV in ninelX and 3X ranges. ACCURACY: f 1% of full scale on all ranges exclusive of noise and drii. ZERO DRIFT: Less than Im” per 24 hours, Ikss than 150 per OC. METER NOISE: *25 *t4% of full scale on 3 x lo-l2 fo 10-14A rano
10、es. METER NOISE: Less than f3 x 10-16A. OFFSET CVRRENT: Lens than 5 x 10-15A. AS AN OHMMETER: RANQE: 7000, full scale to 10% in twent+hree linear IX and 3X ranges. ACCURACY: *3% of full scale on 100 to lQ ranges using the largest available multiplier stting; f5% of full scale on 3 x 109 to 1Ojn rang
11、es. AS AN COULOMBMETER: RANGE: lo-? full scale to 10dC in fifteen IX and 3X ranges. ACCURACY: f5% of full scale on all ranges. Dritt due fo offset current does no, exceed 5 x 10.=C per second. AS AN AMPLIFIER: tNPUT IMPEDANCE: Greater than 1040 shunted by2OpF. input resistance may a,so be nanual. 1.
12、 The Model 602 has excellent zero stability which permits The symbol # on the instrument indicates that a po- accurate measurements with minimal adjustment. Short tential of 1OOOV or more may be present on the terminaltsl. term zero drift is less than 50he NOTE terminal voltage. If the leakage curre
13、nt is less than the offset current (less than 5 X 10.5A there may be no decay of the Using the center zero scales decreases accuracy terminal voltage when the Model 602 is left connected across 0.5% because the scale span is reduced. the capacitors terminals. 2. After a few moments increase the volt
14、age sensitivity by NOTE Keep the shield cap on the INPUT connector when the Model 602 is not in use. advancing the multiplier switch. Continue zeroing with the FINE ZERO control. 3. After long periods of storage or after an overload, the Model 602 may drift excessively. The input transistors are The
15、 Model 6012 Triaxial-to-Coaxial Adapter permits use of cables and accessories with the Model 602 by adapting the triaxial INPUT connector to the UHF coaxial type. insensitive to mechanical shock; however, a severe input overload may cause a zero offset. This is corrected with the ZERO controls. Drif
16、ting can occur for several hours. NOTE CAUTION The Model 6012 connects low to case ground. The Modal 602 cannot be used off- ground when using the Model 6012. The in- struments cabinet will be at the same If the Model 602 has been stored for some time, the offset current will exceed the specificatio
17、n when first used, then decrease to below the specified amount after one or two hours of use. This is an inherent characteristic of the input transistors; the instrument is not faulty. potential as the input low. 4. Although the offset current of the Model 602 is below that 2.5 PRELIMINARY PROCEDURE
18、S found in conventional voltmeters, it can be observed on the meter. The current charges the input capacitance, and Check battery condition by holding the METER switch in the the Model 602 appears to drift when the input is open. Use BATTERY CHECK position. Rotate the multiplier switch the ZERO CHEC
19、K button to discharge the charge build-up. through the ,001 to 0.1 positions and observe the meter Depressing the ZERO CHECK button removes all signal readings. The meter should read at least two-thirds of full from the amplifier and reduces the input impedance scale in each multiplier switch positi
20、on. Table 2-2 shows the between HI and LO terminals of the INPUT connector to 10MR. 2-l 5. Follow the procedures in paragraphs 2.6 to 2.10 for measuring voltage, current, resistance and charge. NOTE When using multiplier switch settings of 10, 3 and 1 in the voltage, current, resistance and charge m
21、easuring modes, make sure the OUT- PUT switch is set to 1V if the output is con- nected to a load of more than 2kR. Otherwise, the meter will not read full scale signals correct- ly. When the output is loaded, this effect is not present. 2.6 VOLTAGE MEASUREMENTS The Model 602 can be used to measure
22、voltages. In the nor- mal method (FEEDBACK switch at NORMAL) the unknown voltage is connected to the INPUT connector. Input im- pedance with the range switch in the VOLTS position is 10%. 2OpF. To reduce the effects of input cable capacity, use the fast method to measure the voltage. Set the FEEDBAC
23、K switch at FAST and drive the inner shield of the cable with the Xl OUTPUT which is connected to the LO terminal in the FAST position, A guarded circuit is possible this way. To reduce stray pick-up when making routine measurements fix. measuring from low impedance sources1 the range resistors or c
24、apacitors may be used to shunt the input. Accessory probes extend the Model 602s range to 1OkV. NOTE Locking the ZERO CHECK switch places 1OMB between input high and low, which may tem- porarily cause instability in e.ome types of high impedance sources. 2.6.1 Normal Method Voltage Measurements Set