Kenwood VT-181_VT-181E 电路图.pdf
《Kenwood VT-181_VT-181E 电路图.pdf》由会员分享,可在线阅读,更多相关《Kenwood VT-181_VT-181E 电路图.pdf(22页珍藏版)》请在收音机爱好者资料库上搜索。
1、AC VOLTMETER VT-181/VT-181E SERVICE MANUAL KENWOOD TMI CORPORATION 1998-4/B51-1132-00 (K/K) HANDLE: (K01-0564-08) HANDLE COVER : (B09-0410-08) FRONT PANEL VT-181: (A63-0301-08) VT-181E : (A63-0316-08) SIDE FRAME (A13-2254-08) PUSH KNOB (K24-3015-08) KNOB (K21-0959-08) RORARY KNOB (K21-0961-08) RUBBE
2、R FOOT (J02-0543-08) VT-181/VT-181E(K) INT 98.4.9 10:26 AM yW 2 RadioFans.CN 收音机爱 好者资料库 VT-181/VT-181E 2 The following instructions are for use by qualified personnel only. To avoid electric shock, do not perform any servicing other than contained in the operating instructions unless you are qualifi
3、ed to do so. WARNING CONTENTS SPECIFICATIONS .3 SAFETY .4 CIRCUIT DESCRIPTION.5 BLOCK DIAGRAM .6 ADJUSTMENT.7 PARTS LIST.9 PARTS LIST (ELECTRICAL).10 SCHEMATIC DIAGRAM .15 P.C. BOARD .21 VT-181/VT-181E(K) INT 98.4.9 10:26 AM yW 3 RadioFans.CN 收音机爱 好者资料库 VT-181/VT-181E 3 SPECIFICATIONS Measurable Vol
4、tage dB dBm Error Frequency response Input impedance Durable input voltage Gain Output voltage Output resistance Frequency response Distortion Signal to noise ratio 1mV to 300mV in 12 ranges:1,3, 10,30,100,300mV,1,3,10,30, 100,300V full scale. -80 to +50dB(0dB=1V) -80 to +52dBm(0dBm=1mW at 600) 500V
5、 (DC +AC peak) 1V to 300V range 100V (DC +AC peak) 1mV to 300mV range Within specifications temp./ hum.range Full operation temp./ hum.range Line voltage Power consumption Dimensions WXHXD(mm) Net Weight Accessories Power cable Input cable Replacement fuse Instruction manual Adjust driver Less than
6、1% at full scale (Rated by signal-noise ratio in 1mV and 1V range) Over 40dB at full scale Within 3% of full scale at 1kHz 0.3mV to 100mV in 12 ranges:0.3, 1,3,10,30,100,300mV,1,3,10, 30,100V full scale. -90 to +40dB(0dB=1V) -90 to +42dBm(0dBm=1mW at 600) 500V (DC +AC peak) 1V to 300mV range 100V (D
7、C +AC peak) 0.3mV to 300mV range Less than 1% at full scale (Rated by signal- noise ratio in 0.3mV,1mV and 1V ranges) Over 30dB at 0.3mV range Power Supply Section Meter Section VT-181EVT-181Item Environmental Indoor Use Only Altitude up to 2000m OVERVOLTAGE CATEGORYPOLLUTION DEGREE2 Amplifier Secti
8、on 10% at 5Hz to 1Mhz, 5% at 10Hz to 500kHz and 2% at 20Hz to 100kHz as reference to 1kHz response. 10M 5%, with less than 50pF parallel capacitance. 60dB 1dB(Approx.70dB) 1Vrms 20% 60020% at 1kHz Within3dB at 5Hz to 500kHz 10 to 40C / 80% RH or less 0 to 50C / 80% RH or less 100/120/220/230/ Vac 10
9、% 50/60Hz Max.6.9W 128(128)X190(210)X239(259) Value in ( ) include protrusions 2.8kg 1pc CA-41p 1pc 1pc 1copy 1pc Stability Residual voltageLess than 20uV with input shorted on 1mV range Less than 30uV with input shorted on 0.3mV range Within 0.5% of full scale for 10 line voltage fluctuation VT-181
10、/VT-181E(K) INT 98.4.9 10:31 AM yW 6 RadioFans.CN 收音机爱 好者资料库 VT-181/VT-181E 4 SAFETY Before connecting the instrument to a power source, care- fully read the following information, then verify that the proper power cord is used and the proper line fuse is installed for power source. The specified vo
11、ltage is shown on the rear panel. If the power cord is not applied for specified voltage, there is always a certain amount of dan- ger from electric shock. Line voltage This instrument operates using ac-power input voltages that 100/120/220/230 V at frequencies from 50 Hz to 60Hz. Power cord The gro
12、und wire of the 3-wire ac power plug places the chassis and housing of the instrument at earth ground. Do not attempt to defeat the ground wire connection or float the instrument ; to do so may pose a great safety hazard. The appropriate power cord is supplied by an option that is specified when the
13、 instrument is ordered. The optional power cords are shown as follows in Fig.1 Line fuse The fuse holder is located on the rear panel and contains the line fuse. Verify that the proper fuse is installed by replacing the line fuse. Voltage conversion This instrument may be operated from either a 100
14、to 230V, 50/60Hz power source. Use the following procedure to change from 100 to 230V operation or vice verse. 1. Remove the fuse holder. 2. Replace fuse F1 with a fuse of appropriate value. 3. Reinsert it for appropriate voltage range. 4. When performing the reinsertion of fuse holder for the volta
15、ge conversion, the appropriate power cord should be used. (See fig.1) SAFETY North American 120 volt/60 Hz Rated 15 amp (12 amp max ; NEC) Universal Europe 230 volt/50 Hz Rated 16 amp U.K. 230 volt/50 Hz Rated 5 amp Australian 240 volt/50 Hz Rated 10 amp North American 240 volt/60 Hz Rated 15 amp (1
16、2 amp max ; NEC) Switzerland 230 volt/50Hz Rated 10 amp Fig.1 Power Input Voltage Configuration Plug configurationpower cord and plug type Factory installed instrument fuse 0.2A, 250V slow blow 5x20mm 0.1A, 250V slow blow 5x20 mm 0.1A, 250V slow blow 5x20 mm 0.1A, 250V slow blow 5x20 mm 0.2A, 250V s
17、low blow 5x20mm 0.3A, 250V slow blow 5x20 mm Line cord plug fuse None None None Parts No. for power cord E30-1982-08 E30-1985-08 E30-1986-08 None 5A Type C None E30-1983-08 VT-181/VT-181E(K) INT 98.4.9 10:31 AM yW 7 VT-181/VT-181E 5 General A Signal voltage to be measured, which is input from the IN
18、PUT connector, is passed through the First Attenuator and is converted to a low impedance by the Impedance Convertor. The impedance-converted signal is normalized, or further attenuated in proportion to 1mVrms fullscale value through the Second and Third Attenuator. The normalized signal is magnifie
19、d 20-fold by the Main Amplifier and is fed to the Output Amplifier and the Absolute-Mean Value Detec- tor. The Output Amplifier magnifies the signal 50-fold and feeds to the OUTPUT connector. The Absolute-Mean Value Detector converts the signal from the Main Amplifier to DC current in proportion to
20、the absolute mean value. The con- verted signal activates the Meter. The Attenuator Control encodes the signal led from the RANGE selector to generate an Attenuator Control signal. This signal controls the First, Second and Third attenuator to set the sensitivity corresponding to each range. The Pow
21、er Supply feeds to the functional circuit 5V DC voltages stabilized by its IC regulator. Description of Functional Circuit 1) First Attenuator A potential divider act as a attenuator. The amount of atten- uation is switched in tow steps by relay contacts:0dB and - 60dB. 2) Impedance Converter A FET
22、differential input Amplifier act as a impedance con- verter with 0dB(10dB) gain. Which converts the First Attenu- ator output signal to a sufficiently low impedance and feeds of the Second Attenuator. 3) Second Attenuator A resistance divider act as a attenuator. The amount of attenuation is switch
23、in two steps by relay contacts:0dB and -30dB. 4) Third Attenuator A resistance divider network act as a attenuator. The amount of attenuation is switched in four steps by FET switch:0dB,-10dB,-20dB,and -30dB. 5) Main Amplifier A wideband,non-inverting differential amplifier act as the main amplifier