1、AP034-OM-E Rev DISSUED: January 2000? ? ? CLEANING The exterior of the probe and cable should be cleaned only using a soft cloth moistened with water or isopropyl alcohol. Abrasive agents, strong detergents, or other solvents may damage the probe. Always ensure that the input receptacles are free of
2、 debris before connecting accessories. CALIBRATION INTERVAL To guarantee accurate performance, you should have the probe calibrated every 12 months. Avoid exposing the probe to extreme mechanical shock and excessive bending of the cable because these may alter the calibration. SERVICE STRATEGY The A
3、P034 circuits utilize fine pitch surface mount devices; it is, therefore, impractical to attempt repair in the field. Defective probes must be returned to a LeCroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe
4、 that is not under warranty can be exchanged for a factory refurbished probe. A modest fee is charged for this service. The defective probe must be returned in order to receive credit for the probe core. TROUBLESHOOTING If the probe is not operating properly, the problem may be the way in which it i
5、s used. Before assuming the probe is defective, perform the following troubleshooting procedures. A. Trace Off Scale This is typically caused by improper offset setting, or by an input signal that exceeds the probe differential or common mode range. Perform the following: 1. Remove the input signal
6、from the probe, return the offset to zero, and Autobalance the probe. Does the trace return to approximately the center of the graticule? If not, proceed to step 7. ! RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? 2. Set the oscilloscope calibrator to output a 100-mV, 1-kHz square w
7、ave. Using the flex lead set, connect the probe + input to the calibrator output signal, leave the input open. Set the oscilloscope to 50 mV/div and 500 s/div. Is the displayed waveform a 100-mV, 1-kHz square wave with the correct polarity? 3. Repeat step 2 with the input connected to the calibrator
8、 and the + input left open. In this case, the displayed waveform should be inverted. 4. Connect both the + input and input to the calibrator output. Is the trace approximately a flat line near zero volts? 5. If steps 1 to 4 give the correct results, the problem is likely a result of the input signal
9、 exceeding the differential or common mode range. 6. Connect both the + input and input to one of the two input signals. If the trace is off scale, the input signal is probably exceeding the common mode range. Repeat with the other input signal. 7. Is a ProBus Power Supply Overload error message dis
10、played? If so, remove all other ProBus accessories from the oscilloscope. Is the message still displayed? If so, remove the AP034. Is the message still displayed? If so, the oscilloscope should be returned for service. 8. If, after removing the AP034, the Power Supply Overload error message is not d
11、isplayed, the problem may be either the probe or the oscilloscope. Repeat the test with a different ProBus accessory. If the message does not return with a different accessory, the AP034 may be defective and should be returned for service. B. Incorrect Frequency Response Possible causes are a defect
12、ive probe or oscilloscope, poor connections, or poor grounding. Try the following: 1. Verify that the BW limiting of the oscilloscope is off. RadioFans.CN 收音机爱 好者资料库 AP034-OM-E Rev DISSUED: January 2000? ? 2. Connect the probe to another oscilloscope. If the probe now measures properly, the problem
13、may be in the oscilloscope. 3. If the probe behaves as if it is ac-coupled at high frequency, check for an open input connection. 4. Poor frequency or transient response and AC gain errors may result when one of the two input connections is open. 5. Excessive “ring” and other transient problems can
14、result from excessive lead length. To test this, shorten the input leads to less than 1 cm. If the transient response changes significantly, the lead parasitics are the cause. C. DC Errors Incorrect DC gain requires recalibration or factory repair. This can be determined by completing the gain check
15、s in the Performance Verification Procedure. 1. Extremely high source resistance will result in DC gain errors. Check the probe accuracy with the oscilloscope calibrator signal. 2. Verify that the probe is not being overdriven into clipping for its current gain setting. 3. Excessive offset can resul
16、t from large changes in ambient temperature. Remove the input signal from the probe and repeat the Autobalance cycle. With the Offset set to zero, did the trace return to the center of the graticule? D. Poor Common Mode Rejection Use the 1-kHz calibrator signal from the oscilloscope to check common
17、mode rejection. With both the + input and input connected to the calibrator signal, a flat line at zero volts should be seen on the graticule. 1. Check the probe with the plug-on attenuator installed and removed. If excessive common mode signal appears only when the attenuator is present, the attenu
18、ator may need to be rematched to the probe. Use the procedure listed in this section to match the attenuator. 2. If the common mode signal appears when the probe is connected to the test circuit, but not when it is attached to RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? the calib
19、rator, the problem may be caused by large mismatches in the source impedance. Try connecting both inputs to one of the input signals in the test circuit, then the other. If the common mode signal disappears, try probing lower impedance points within the circuit. RETURNING A DEFECTIVE PROBE The proce
20、dure for returning a defective probe to be exchanged, is as follows: Contact your local LeCroy sales representative to find out where to return the product. All returned products should be identified by model and serial number. You should describe the defect or failure, and provide your name and con
21、tact number. In the case of products returned to the factory, a Return Authorization Number (RAN) should be used. The RAN can be established by contacting your nearest LeCroy office, or the New York Customer Care Center. Return shipments should be made prepaid. LeCroy cannot accept COD or Collect Re
22、turn shipments. We recommend air- freighting. It is important that the RAN be clearly shown on the outside of the shipping package for prompt redirection to the appropriate LeCroy department. 1. Contact your local LeCroy sales or service representative to obtain a Return Authorization Number. 2. Rem
23、ove all accessories, including all removable cables, from the probe. 3. Pack the probe in its case, surrounded by the original packing material (or equivalent) and box. 4. Label the case with a tag containing: The RAN Name and address of owner Instrument model and serial number Description of the fa
24、ilure mode 5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit. AP034-OM-E Rev DISSUED: January 2000? ? 6. Mark the outside of the box with the shipping address given to you by the LeCroy representative; be sure to add the following: ATTN: FRAGILE 7