lecroy AP034-OM-E-06 电路图.pdf
《lecroy AP034-OM-E-06 电路图.pdf》由会员分享,可在线阅读,更多相关《lecroy AP034-OM-E-06 电路图.pdf(12页珍藏版)》请在收音机爱好者资料库上搜索。
1、AP034-OM-E Rev DISSUED: January 2000? ? ? This procedure can be used to verify the warranted characteristics of the AP034 Active Differential Probe. The recommended calibration interval for the model AP034 Active Differential Probe is one year. The complete performance verification procedure should
2、be performed as the first step of annual calibration. You can record test results on a photocopy of the Test Record provided at the end of this section. You can do the performance verification without removing the instrument covers, exposing yourself to hazardous voltages. Adjustment should only be
3、attempted if a parameter measured in the Performance Verification Procedure is outside of the specified limits. Adjustment should only be performed by qualified personnel. TEST EQUIPMENT REQUIRED Table 3 lists the test equipment and accessories (or their equivalents) that are required to verify the
4、performance of the AP034 Active Differential Probe. This procedure is designed to minimize the number of calibrated test instruments required. Only the parameters listed in boldface in the “Minimum Requirements” column must be calibrated to the accuracy indicated. Because the input and output connec
5、tor types may vary on different brands and models of test instruments, additional adapters or cables may be required. RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? TABLE 3. List of Required Equipment DescriptionMinimum RequirementsTest Equipment Examples Wide Band Oscilloscope 1Min
6、imum 1 GHz bandwidth 2 mV to 5 V scale factors ProBus interface equipped 2% vertical accuracy LeCroy LC584A Digital MultimeterDC: 0.1% accuracy AC: 0.2% accuracy to measure 200 mV and 2 V rms 1 kHz 5 digit resolution Agilent Technologies 34401A Fluke 8842A-09 Keithley 2001 Function GeneratorSine Wav
7、e and Square Wave output waveforms 20 Vp-p? 70 Hz to 10 MHz frequency range Agilent Technologies 33120A Stanford Research Model DS340 Leader LAG-120B High Frequency Sine Wave Generator 50 to 500 MHz and 50 kHz Output adjustable from 300 mV to 2 Vp-p Agilent Technologies 8648C Terminator, in-line, BN
8、C? ?Pomona 4119-50 Terminator, precision, BNC? 0.2%LeCroy TERM-CF01 Attenuator, BNC? ?10 (20 dB)Pomona 4108-20dB BNC coaxial cable (2 required) ? ?Pomona 5697-36 Calibration FixtureProBus Extension CableLeCroy PROBUS-CF01 Calibration FixtureAP033/AP034 Calibration FixtureLeCroy AP03x-CF01 Banana Plu
9、g adapterBNC female-to-banana plugPomona 1269 Notes 1 If a LeCroy ProBus equipped oscilloscope is not available, you may use an alternate oscilloscope that meets the other minimum specifications listed, and the model ADPPS power supply, to perform the adjustment procedure. The input RadioFans.CN 收音机
10、爱 好者资料库 AP034-OM-E Rev DISSUED: January 2000? ? ? ?!? offset or position must remain at center screen. PRELIMINARY PROCEDURE 1. Connect the AP034 Active Differential Probe to the female end of the ProBus Extension Cable. Connect the male end of the ProBus extension cable to Channel 1 of the oscillos
11、cope. 2. Turn the oscilloscope on and allow at least a 30- minute warm-up time for the AP034 and test equipment before performing the Verification Procedure. 3. Turn on the other test equipment and allow these to warm up for the time period recommended by the manufacturer. 4. While the instruments a
12、re reaching operating temperature, make a photocopy of the Performance Verification Test Record (located at the end of this section), and fill in the necessary information. The warranted characteristics of the AP034 Active Differential Probe are valid at any temperature within the Environmental Char
13、acteristics listed in the Specifications. However, some of the other test equipment used to verify the performance may have environmental limitations required to meet the accuracy needed for the procedure. Make sure that the ambient conditions meet the requirements of all the test instruments used i
14、n this procedure. PROCEDURE Note Correct operation of the AP034 controls requires oscilloscope software version 8.1.0 or higher. The use of earlier versions is not recommended. You can verify the software version in the test oscilloscope by pressing the SHOW STATUS button, then selecting the System
15、menu option. Contact your local LeCroy representative if the software in your oscilloscope requires updating. RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? A. Check Gain Accuracy 1. From the oscilloscope display, select the channel that the AP034 is connected to (channel 1), then s
16、elect the “Coupling” menu. Set the Coupling to DC 1 M. (If the only choice available is AC 1 M, remove the Plug-on AC Coupling adapter from the probe tip.) 2. If necessary, set the probe offset to 0.000 V by rotating the OFFSET knob in the CHANNEL section of the oscilloscope. 3. Using a BNC female-t
17、o-female adapter, connect one end of a BNC cable to the probe end of the ProBus Extension ?!?#?$? end of the BNC cable. 4. Connect one end of a second BNC cable to the output of the Function Generator. Attach the BNC-to-dual-male banana plug adapter to the free end of the second BNC cable. Connect t
18、he banana plugs to the Digital Multimeter (DMM). 5. Set the DMM to measure AC Volts. 6. Set the mode of the Function Generator to Sine Wave; the frequency to approximately 1 kHz; and the output amplitude to approximately 200 mV rms, as measured by the DMM. 7. Record the measured amplitude to 100 V r
19、esolution in the Test Record. 8. Unplug the output cable from the DMM. Remove the BNC-to- banana plug adapter from the end of the cable. 9. Autobalance the AP034 by pressing the AUTOBALANCE menu button located in the “Coupling” menu. 10. Carefully align the four pins that correspond to the Different
20、ial Drive No Termination portion of the AP03x- CF01 Calibration Fixture with the input receptacles in the AP034 probe head. Press the probe into the fixture to fully engage the pins. 11. Connect the banana plugs of the precision terminator to the digital multimeter (DMM). AP034-OM-E Rev DISSUED: Jan
21、uary 2000? ? 12. Connect the free end of the output cable from the Function Generator to the Differential Drive No Termination connector of the AP033/AP034 Calibration Fixture. 13. After the DMM reading has stabilized, record the measured output amplitude to 100 V resolution in the Test Record. 14.
22、Divide the measured output voltage (recorded in step A-13) by the sine wave generator output voltage (probe input voltage) from step A-7. Subtract 1.0 from the ratio and multiply the result by 100 to get the error in percent. 1001% = VoltageInput ageutput VoltMeasured O Error 15. Record the answer t
23、o two significant places (x.xx%) on line A-15 in the Test Record. 16. Verify that the X1 gain error is less than 2%. 17. Disconnect both BNC cables from the test setup. B. Check High Frequency Common Mode Rejection Ratio (CMRR) Note Common Mode Rejection Ratio (CMRR) is defined as the Differential M
24、ode Gain divided by the Common Mode Gain (normalized inverse of the Common Mode response). At frequencies below approximately 250 MHz, the bandwidth- related attenuation in the AP034 Active Differential Probe is so insignificant that the Differential Mode Gain can be assumed to be unity (1.0). 1. Di