1、AP034-OM-E Rev DISSUED: January 2000? ? ? INTRODUCTION You can use this procedure to adjust the AP034 Active Differential Probe to meet the warranted specifications. This procedure should only be performed if the instrument fails the Performance Verification tests. If the probe cannot be adjusted to
2、 meet the Performance Verification limits, repair may be necessary. To ensure instrument accuracy, check the calibration of the AP034 Active Differential Probe every year. Before calibration, thoroughly clean and inspect this unit as discussed in the “Cleaning” section. Completion of each step in th
3、e Adjustment Procedure ensures that the differential probe meets specifications. Some of the adjustments interact with other parts of the circuitry. Therefore, it is necessary that all adjustments be performed in the order listed. For best overall instrument performance, make each adjustment to the
4、exact setting, even when adjustment is within the limits stated in the procedure. Adequate guard bands were designed into the AP034 Active Differential Probe to ensure that it will meet or exceed published specifications over the entire operating temperature range. To continue to meet the environmen
5、tal specifications, all adjustments must be performed in a controlled environment with an ambient temperature of 25 5 ? ? ? ? Differential Probe must also be at stable operating temperature before performing adjustments. Caution The adjustment procedure will require removal of the probe covers. Thes
6、e covers are part of the ESD protection system of the AP034 Active Differential Probe. To protect the probe, you should perform the entire procedure on a static dissipating work surface. Wear an antistatic RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-OM-E Rev D ? grounding wrist strap and foll
7、ow standard static control procedures. The probe tip housing provides physical rigidity to the input pins of the probe. When the covers are removed, observe extra caution to avoid breaking the probe tip receptacles when mating the probe to the calibration fixture. TEST EQUIPMENT REQUIRED The table o
8、n the next page lists the test equipment and accessories, or their equivalents, that are required for complete calibration. Specifications given for the test equipment are the minimum necessary for accurate calibration. All test equipment is assumed to be correctly calibrated and operating within th
9、e specifications listed. Detailed operating instructions for the test equipment are not given in this procedure. Refer to the test equipment manual if more information is needed. If alternate test equipment is substituted, control settings or calibration equipment setups may need to be altered. Alte
10、rnate models of test equipment may have different connector styles requiring adapters not included in the equipment list. RadioFans.CN 收音机爱 好者资料库 AP034-OM-E Rev DISSUED: January 2000? ? TABLE 4. Test Equipment and Accessories DescriptionMinimum SpecificationsTest Equipment Examples Wide Band Oscillo
11、scope * 500 MHz bandwidth 2 mV to 200 mV vertical scale factors ProBus interface equipped* 2% vertical accuracy LeCroy LT342 LeCroy LC344A Digital Multimeter0.1% DC volts accuracy at 0.4 V 0.2% AC volts accuracy at 2 V and 1 kHz 5 digit resolution 0.01 mV AC volts resolution Agilent Technologies 344
12、01A Fluke 8842A-09 Keithley 2001 with test leads Function GeneratorSine and square wave output 20 Vp-p? 50 Hz to 1 MHz frequency range Sync. or Freq. Reference output Agilent Technologies 33120A Stanford Research Model DS340 Leader LAG-120B Fast Rise Pulse Generator Risetime 2 ns ? Over/Undershoot 6
13、% Calibrator signal from LeCroy oscilloscope Tegam / Tektronix PG506 with power unit Sine Wave GeneratorOutput 250 MHz at 1 V rmsAgilent Technologies 8648C, with male N to female BNC adapter for the output connector Tegam SG504 with TM series mainframe. DC Voltage Source0 to 2 VDC regulated output,
14、settable to 1 mV resolution HP 6633A Termination, precision? ?LeCroy TERM-CF01 Termination, BNC? ? ?!?#$?Pomona 4119-50 Calibration FixtureProBus Extension CableLeCroy PROBUS-CF01 Calibration FixtureAP033/AP034 Calibration Fixture LeCroy AP03X-CF01 RadioFans.CN 收音机爱 好者资料库 ?ISSUED: January 2000AP034-
15、OM-E Rev D ? TABLE 4. Test Equipment and Accessories DescriptionMinimum SpecificationsTest Equipment Examples BNC coaxial cable (2 required) male-male BNC, approx. 1 mPomona 5697-36 BNC Tee Adapterfemale-male-femalePomona 3285 Banana Plug Adapter (2 required) BNC female-to-dual male banana plug Pomo
16、na 1269 Shorting PlugTwo 0.025 in. square pins on 0.100 in. centers Supplied with AP03X-CF01 Adjustment Tool0.025 in. square tipJohanson 4192 Low Capacitance Screwdriver * Notes If a LeCroy ProBus equipped oscilloscope is not available, you may perform the adjustment procedure with an alternate osci
17、lloscope that meets the other Minimum Specifications, and the ?%?&?(?)?$?(?(?(?*?(?!? offset or position must remain at center screen, and the high-frequency adjustments must be performed with the ADPPS connected directly to the input of the oscilloscope. PRELIMINARY PROCEDURE 1. Using pliers, caref
18、ully remove each of the push buttons from the ProBus interface housing by gently pulling them away from the housing. 2. Remove the two screws that secure the plastic cover on the cable end of the ProBus interface housing. Gently pull on the probe cable to slide the circuit board assembly from the me
19、tal housing. 3. Remove the 5/64 in. (2 mm) Allen head cap screw from the small cover on the back of the probe tip. Remove the cover along with the small cover on the opposite side of the probe. Hold the shielded portion of the probe head in one hand and AP034-OM-E Rev DISSUED: January 2000? ? gently
20、 slide the larger cover off by pulling it away from the probe tip end. 4. Connect the AP034 Active Differential Probe output to the female end of the ProBus Extension Cable. Be careful to align the ProBus pins with the corresponding connector correctly. Connect the male end of the ProBus Extension C
21、able to Channel 1 of the oscilloscope. Note The Logic board is connected to the Amplifier board by four small 8-pin connectors. There are no additional mechanical fasteners holding the two boards together. A small foam pad mounted on the circuit board ensures that the connectors remain engaged while
22、 the assembly is mounted within the ProBus interface housing. Be sure that the connectors between the boards are firmly engaged before applying power to the probe. Operating the probe without the logic board will not damage it. However, to ensure reliable operation, the logic board should only be ma
23、ted with the Amplifier board with the power removed, and the ProBus connector disengaged. 5. Apply power to the oscilloscope and to the other test instrumentation. Allow at least 30 minutes warm-up time for the AP034 Active Differential Probe and test equipment before continuing the calibration proc
24、edure. Note The probe tip cover also serves to thermally stabilize the input circuitry. The differential input stage of the probe utilizes discrete transistors that need to maintain an approximate match of junction temperatures for correct DC balance. With the covers removed, this circuitry is susce