1、MANUAL INSTRUCTION RadioFans.CN 收音机爱 好者资料库 NOTES AND CAUTIONS ELECTRICAL SAFETY PRECAUTIONS This equipment is protected in accordance with IEC Safety Class 1. It has been designed and tested according to IEC Publication 348, Safety Requirements for Electronic Measuring Apparatus, and has been suppli
2、ed in a safe condition. The following pre- cautions must be observed by the user to ensure safe operation and to retain the equip- ment in a safe condition. Defects and abnormal stresses Whenever it is likely that protection has been impaired, for example as a result of damage caused by severe condi
3、tions of transport or storage, the equipment shall be made inoperative and be secured against any unintended operation. Removal of covers Removal of the covers is likely to expose live parts although reasonable precautions have been taken in the design of the equipment to shield such parts. The equi
4、pment shall be disconnected from the supply before carrying out any adjustment, replacement or maintenance and repair during which the equipment shall be opened. If any adjustment, maintenance or repair under voltage is inevitable it shall only be carried out by a skilled person who is aware of the
5、hazard involved. Note that capacitors inside the equipment may still be charged when the equipment has been disconnected from the supply. Before carrying out any work inside the equip- ment, capacitors connected to high voltage points should be discharged; to discharge mains filter capacitors, if fi
6、tted, short together the L (live) and N (neutral) pins of the mains plug. Mains plug The mains plug shall only be inserted in a socket outlet provided with a protective earth contacts The protective action shall not be negated by the use of an extension lead without protective conductor. Any interru
7、ption of the protective conductor inside or outside the equipment is likely to make the equipment dangerous. Fuses Note that there is a supply fuse in both the live and neutral wires of the supply lead. If only one of these fuses should rupture, certain parts of the equipment could remain at supply
8、potential. To provide protection against breakdown of the supply lead, its connectors, and filter where fitted, an external supply fuse (e.g. fitted in the connecting plug) should be used in the live lead. The fuse should have a continuous rating not exceeding 6 A. Make sure that only fuses with the
9、 required rated current and of the specified type are used for replacement. The use of mended fuses and the short-circuiting of fuse holders shall be avoided. RADIO FREQUENCY INTERFERENCE This equipment conforms with the requirements of lEC Directive 76/889 as to limits of r.f. interference. (2a) Ra
10、dioFans.CN 收音机爱 好者资料库 Contents Section 1 GENERAL INFORMATION 1.1 1.2 1.3 Section 2 OPERATION 2. 2. 2. 2. 2. 2. 2. 2.8 2.9 2.10 2.11 ,1 ,2 .3 .4 5 ,6 .7 Introduction Data summary Accessories Installation Controls and connectors Preliminaries Connecting the component Resistance measurement Inductance
11、measurement Notes on inductlve loss balancing Using the D. C. Choke Adaptor Type TM Capacitance measurement Notes on capacitive loss balancing Use of external oscillator and detector 6113 9 10 10 12 16 17 18 21 21 Section 3 TECHNICAL DESCRIPTION 3.1 Circuit summary 3.2 Measuring bridge a. c. supply
12、3.3 Measuring bridge d. c. supply 3.4 Measuring bridge circuits 3.5 Amplifier detector 3.6 Power supplies 3.7 D. C. Choke Adaptor Type TM 6113 23 24 24 24 25 26 26 Section 4 MAINTENANCE 4.1 General 4. 2 Removal from case 4.3 Access to components 4.4 Mains input adjustments 4.5 Working voltages 4.6 R
13、eplacement of valves 4. 7 Adjustment of preset components 4.8 Checking the loss balance diai calibration 4. 9 Access and adjustment to balance control assembly 28 28 28 28 28 31 31 36 36 Section 5 REPLACEABLE PARTS Universal bridge D. C. choke adaptor 41 53 Section 6 CIRCUIT DIAGRAMS Fig. 6.1 Power
14、supply and oscillator Fig. 6.2 Bridge Fig. 6.3 Detector Fig. 6.4 D. C. choke adaptor 47 49 51 53 1313A (1) RadioFans.CN 收音机爱 好者资料库 General information 1.1 INTRODUCTION TF 1313A is a general-purpose impedance bridge with one tenth per cent measurement accuracy over a wide range of inductance, capac-
15、itance, and resistance values. It offers except- ional discrimination and resettability, wide-range loss balancing, and facilities for using an external oscillator and detector. The single direct-reading dial used for L, C, and R measurements has coarse aiid fine con- centric controls; the coarse on
16、e moves in 1% steps and the fine gives continuously variable interpolation between steps. The variable and non-linear sensitivity of the detector facilitates easy measurement of completely unknown comp- onents. An external audio oscillator can be plugged in where L and C measurements are req- uired
17、at frequencies other than the internal 1 and 10kc/s; the detector output is available externally to allow an oscilloscope or headphones to be used for balance indication. This bridge can be used for precision eval- uation of resistance, capacitance and inductance, for measuring the dissipation facto
18、r of capacitors and Q of inductors, and for quickly identifying unknown components. Measurements on high-loss components are made easier by the low D and Q ranges of the loss balance control and the relative independence between the adjustment of the main and loss balance controls. Its high discrimi
19、nation and resettability make it particularly suitable for comparative measurements such as checking the difference between an unknown and a standard component. Measurements on inductors carrying d. c. can be made by using Adaptor TypeTM6113, which is available as an optional accessory, while an ext
20、ernal d. c. supply can be connected direct for polarizing capacitors. Fig. 1.1 0-1% Universal bridge TF 1313A 1313A (la) 1.2 DATA SUMMARY Resistance measurement General information Range Accuracy : Residual resistance 3 m to 110 M. in eight ranges with maxima of 11, 110, 1.1 k, 11 k, 110 k, 1.1 M, 1
21、1 M, 110 M. Basic measurement error : 0.1% of reading, or 0. 015% of rang maximum, whichever is greater. Rang errors : 110 to 1.1 M ranges inclusive - basic errors only. 11 and 11 M ranges - basic errors, and additional 0.1% of reading. 11 M X 10 rang - basic errors, and additional0.15% of reading.
22、Less than 0.003 . Inductance measurement Rang : Accuracy : 0.1 H to 110 H in seven ranges, with maxima of 110 H, 1.1 mH, 11 mH, 110 mH, 1.1 H, 11 H, 110 H. Basic measurement error at 1 kc/s : 0.1% of reading, or 0. 015% of rang maximum whichever is greater. Basic measurement error at 10 kc/s : 0. 2%
23、 of reading, or 0. 025% of range maximum, whichever is greater. Residual inductance : Range errors : 1.1 mH to 11 H ranges inclusive - basic errors only. 110 H and 110 H ranges - basic errors, and additional 0.1% of reading. Additional errors at Low Q : (0.1 xf/Q)%, where f is in kc/s. Typically 0.
24、05 H. Capacitance measurement Range : Accuracy : (when D is not greater than 0.031) 0.1 pF to 110 F in seven ranges, with maxima of 110 pF, 1100 pF, 0.011 F, 0.11 F, 1.1 F, 11 F, 110 F. Basic measurement error at 1 kc/s : 0.1% of reading, or 0. 015% of range maximum, whichever is greater. (When D is
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- Marconi_TF-1313A 电路图 Marconi_TF 1313
收音机爱好者资料库所有资源均是用户自行上传分享,仅供网友学习交流,未经上传用户书面授权,请勿作他用。