1、 Second Edition January 2008 Audio Analyzers UP 300/UP 350 10 Hz to 80 kHz The new product family from Rohde can also be superimposed on the sinewave signals Audio Analyzers UP 300/UP 35012 5 Diverse high-quality test signals The generators in the Audio Analyzers UP 300 and UP 350 set new stan- dard
2、s in the lower price segment. By providing a wide variety of sinewave signals, two-tone and multitone signals, bursts and noise, the instruments offer the ideal test signals for measurements in the lab, in service and in production, as well as in uni- versity education. Because the audio analyzers h
3、ave inherent distortion of less than 90 dB, even high-end audio devices can be measured. High-end measurement characteristics The audio analyzers offer bandwidths of up to 80 kHz, enabling the user to perform measurements even on broadband audio equipment. The UP 350 is capable of sampling rates up
4、to 192 kHz unprecedented in this class of instruments. Extensive selection of filters The UP 300 and UP 350 contain a wide variety of weighting, third-octave and octave filters. As many as three filters can be combined. Powerful FFT analysis The FFT analysis capability of the UP 300/350 also sets ne
5、w standards in this class of instruments. This capability supports up to 16 k points and provides numer- ous window functions, which enables it to display the spectral composition of sig- nals up to 80 kHz in bandwidth. 6Audio Analyzers UP 300/UP 350 The new instrument family equipped for the future
6、 Versatile applications Desktop use Portable for mobile use Integration into 19-inch racks 7Audio Analyzers UP 300/UP 350 USB interfaces The USB host interface links the instruments to the PC world. The bus ensures high data transmission rates at low cost. Other peripherals (e.g. printers) can be ad
7、dressed via another USB interface. Identical housing All instruments based on the Family 300 concept have an almost identical “face”, a 5.4-inch TFT display, front- panel control elements, protective guards and a handle that can be adjusted to different positions. Only the connectors on the front an
8、d rear panel vary depending on the instrument type. If the protective guards and the handle are removed, the UP 300/350 can be installed in a 19-inch rack. Owing to their slim design, two instruments of the Family 300 can be placed next to each other. In addition to its analog interfaces, the UP 350
9、 has digital BNC interfaces in consumer and professional format on its rear panel. UP 300 UP 350 Audio Analyzers UP 300/UP 3508 Analyzer Analog audio inputs Digital audio inputs (model UP 350 only) Frequency rangeDC/10 Hz to 80 kHz Frequency response (referenced to 1 kHz) 10 Hz to 20 Hz 20 Hz to 22
10、kHz 22 kHz to 40 kHz 40 kHz to 80 kHz 0.1 dB 0.05 dB 0.1 dB 0.25 dB BNC connectors2 channels, floating or grounded (from serial no 100050), selectable AC/DC coupling, channel 1 on front panel, channel 2 on rear panel Maximum input voltagerms, sinewave33 V Measurement rangesin steps of 6 dB390 mV to
11、50 V (max. input 33 V) Input impedanceinner/outer conductor to ground100 k Crosstalk attenuationfrequency 100 dB Common-mode rejectionat 50 Hz, Vin 3 V at 1 kHz, Vin 3 V at 16 kHz, Vin 80 dB 75 dB 60 dB Generator outputeach input channel switchable to the other generator output channel BNC connector
12、unbalanced, grounded, on rear panel Impedance75 Input level (Vpp)100 mV to 5 V Optical inputTOSLINK Channels1, 2, or both Audio bits16 to 24 Sampling rate32 kHz, 44.1 kHz, 48 kHz, 96 kHz, 192 kHz Formatprofessional and consumer Specifications Important: We continously refine our products. Please che
13、ck our homepage www.up300.rohde- for new applications and features. Specifications apply under the following conditions: specified environmental conditions met, calibration cycle adhered to and total calibration performed. 9Audio Analyzers UP 300/UP 350 Measurement functions RMS value, wideband Erro
14、r limitsmeasurement speed AUTO, at 1 kHz sine, AC coupling additional error with measurement speed AUTO FAST additional error with DC coupling 0.1 dB 0.1 dB 0.1 % of measurement range Integration timeAUTO FAST/AUTO VALUE 5 ms/50 ms, at least 1 cycle 1 ms to 10 s Noisewith A filter, 600 source impeda
15、nce with CCIR unweighting filter, 600 source impedance 2 V 4 V Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined RMS value, selective Error limits0.2 dB Bandwidth (0.1 dB)fixed bandwidth filters3 Hz, 10 Hz, 30 Hz, 100 Hz or 300 Hz Bandwidth (3 d
16、B)relative bandwidth absolute bandwidth 1 %, 3 %, 1/12 octave, 1/3 octave, value (10 Hz minimum) 10 Hz to fmax/5 Selectivity100 dB Frequency settingfixed through entered value or automatic to input signal Peak value Measurementpos. peak, neg. peak, peak-to-peak, absolute peak Error limitsat 1 kHz0.2
17、 dB Interval20 ms to 10 s Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined Quasi-peak Measurementin accordance with CCIR 468-4 Error limitsanalyzer bandwidth 22 kHzin accordance with CCIR 468-4 Noisewith CCIR weighting filter, 600 source impeda
18、nce12 V Filtersweighting filters and predefined octave and third-octave filters; up to 3 filters can be combined DC voltage Voltage range0 V to 33 V Error limits (1 % of measured value + 0.5 % of measure- ment range) Total harmonic distortion (THD) Fundamental20 Hz to 20 kHz Frequency tuningfixed th
19、rough entered value, auto-tuning to input signal Weighted harmonicsup to 80 kHzany combination of d2 to d9 Error limitsharmonics 50 kHz harmonics 80 kHz 0.7 dB 1 dB Inherent distortionfundamental 1 kHz fundamental 20 Hz to 5 kHz fundamental 5 kHz to 15 kHz fundamental 15 kHz to 20 kHz 100 dB 90 dB 8
20、5 dB 80 dB Spectrumbargraph showing signal and distortion Audio Analyzers UP300/UP35010 Measurement functions THD+N and SINAD Fundamental20 Hz to 20 kHz Frequency tuningfixed through entered value, auto-tuning to input signal Bandwidthweighting filters and predefined octave and third-octave filters;