HP-3048A_cal_03048_90015 电路图.pdf
《HP-3048A_cal_03048_90015 电路图.pdf》由会员分享,可在线阅读,更多相关《HP-3048A_cal_03048_90015 电路图.pdf(254页珍藏版)》请在收音机爱好者资料库上搜索。
1、Errata Title been impaired, the instrument must be made. inoperative and be secured against any uhin-; tended operation. If this instrument is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to the earth terminal of the power source. Servicin
2、g instructions are for use by service trained personnel only. To avoid dangerous elec tric shock, do not perform any servicing unless qualified to do so. Adjustments described in the manual are per formed with power supplied to the instrument while protective covers are removed. Energy available at
3、may points may, if contacted, result in personal injury. Capacitors inside the instrument may still be charged even if the instrument has been discon nected from its source os supply. For continued protection against fire hazard, re place the line fuse(s) only with 250V fuse(s) of the same current r
4、ating and type (for exam ple, normal blow, time delay, etc.) Do not use repaired fuses or short circuited fuseholders. ATTENTION Static Sensitive Devices This instrument was constructed in an ESD (electro-static discharge) protected environment. This is because most of the semi-conductor devices use
5、d in this instrument are susceptible to damage by static discharge. Depending on the magnitude of the charge, device substrates can be punctured or destroyed by contact or mere proximity of a static charge. The results can cause degradation of device performance, early failure, or immediate destruct
6、ion. These charges are generated in numerous ways such as simple contact, separation of materials, and normal motions of persons working with static sensitive devices. When handling or servicing equipment containing static sensitive devices, adequate precautions must be taken to prevent device damag
7、e or destruction. Only those who are thoroughly familiar with industry accepted techniques for handling static sensitive devices should attempt to service circuitry with these devices. In all instances, measures must be taken to prevent static charge build-up on work surfaces and persons handling th
8、e devices. Model 3048A Table of Contents TABLE OF CONTENTS General Information 1 Quick Check 5 Spur Accuracy Test 7 Noise Flatness Test 11 Noise Floor Test 14 Microwave Phase Detector Conversion Loss (Option 201) Test 16 Functional Tests 19 DAC Tests 21 A3 Adjustments 23 A4 Adjustments 29 Option 1 C
9、alibration 37 Option 2 Calibration 40 Internal Sources Calibration 46 Spectral Purity Tests for Options 001 and 002 49 Spectral Purity Tests for Options 003 and 004 51 Spectral Purity Tests for Options 005 and 006 55 Appendix A: Block Diagram and System Troubleshooting : A-l Appendix B: HP 8662A or
10、HP 8663A 640 MHz Spectral Purity Test B-l Model 3048A Calibration General Information INTRODUCTION This manual documents the procedures which calibrate the HP 3048A Phase Noise Measurement System. System calibration assures that the System meets its published specifications. The proce dures consist
11、of system checks, adjustments, performance tests, and signal path characterization. Calibration of the HP 11848A Phase Noise Measurement Interface results from calibration of the System. NOTE Calibration as referred to in this manual should not be confused with the measurement calibration referred t
12、o in the normal course of making a phase noise measurement. Measurement calibration (especially, a mea surement with phase lock) refers to the characterization of such parameters as the detector constant, tuning sensitivity of the signal sources FM input, and determination of the phase lock loop ban
13、dwidth. iter The need for calibration is governed by the situation. It is usually not necessary nor even desirable to recalibrate the System frequently. Testing guidelines are summarized below. Most tests are automatic. Operator intervention is documented in the Systems test software. It is recommen
14、ded, however, that the user follow along in this manual as the tests are running. EQUIPMENT REQUIRED Equipment required, but not part of the System, is minimal. A feature of the System is the four sources built into the Interface which substitute for external sources in many cases. Extensive use is
15、made of the HP 3561A Dynamic Signal Analyzer both as a signal source (using its built-in noise source) and as a signal analyzer. Any calibrated HP 3561A can be used in calibrating the System; it does not need to be the specific HP 3561A in the System. The external test equipment requirements are giv
16、en in Table 1. PERFORMANCE TEST RECORD Most test results can be hardcopied on the Systems printer. This practice is recommended. It is also recommended that the serial number of the Interface under test be recorded in the System Configuration Table so that it will appear on the printouts. rev. 29MA
17、Y90 General Information Calibration Model 3048A Table 1. Recommended Test Equipment Instrument Type Frequency Counter Microwave Sources Noise Floor Test Fixture Power Meter and Sensor Printer Critical Specifications Frequency Range: 10 to 550 MHz For systems with Option 201 only(1); 2 required Frequ
18、ency Range: 1.2 to 18 GHz Maximum Level: +7 dBm (source 1), 0 dBm (source 2) Stability: drift small compared to 100 kHz Supplied with system. No substitute recommended. For systems with Option 201 only(2) Frequency Range: 1.2 to 18 GHz Level Range: +7 dBm HP-IB; graphics; usually part of the System
19、Suggested Model HP5315A See note (2). HP 11848-61032 HP 435B HP 8481A HP ThinkJet (1) Option 201 adds the 1.2 to 18 GHz input. 2 A variety of combinations of sources with adequate output and frequency range can be used for this test. Table 2. Performance Verification Guideline Situation1) New Instal
20、lation Annual Calibration Ambient Change Confidence Check HP 11848A Repair HP 3561A Repair Check or Test to Run Quick Check X X X Performance Tests xro X3) x X(3) Functional tests X(3) X(3) DAC Tests X(3) X(3) Calibration Option 1 X Option 2 (3) (3) Sources X(3) X(3) W Perform the Adjustments when r
21、ecommended by the other tests. (2) Run the Performance Tests when it is desired to ensure that the System meets its published specifications. (3) Perform these in the following order: Functional Tests, DAC Tests, Option 2 Calibration, Internal Sources Calibration, then Performance Tests. 2 General I
22、nformation Model 3048A Calibration WHEN AND WHAT TO TEST Use Table 2 as a guideline for verifying the performance of the System. The checks and tests are summarized below. The procedures are listed in the order in which they are described except for the Source Options Spectral Purity Tests which fol
23、low the Adjustments. Quick Check is a confidence check which performs a complete phase-locked measurement of the phase noise of the 10 MHz A vs. B internal sources. Performance Tests verify that the System meets its published specifications. The tests are as follows: Spur Accuracy Test verifies the
24、accuracy of noise measurements by measuring the level of phase modulation on a carrier with a known discrete sideband level. (In the context of phase noise measurements, discrete sidebands are often referred to as spurious signals or spurs.) Noise Flatness Test measures the unflatness of noise signa