1、HP Archive This vintage Hewlett Packard document was preserved and distributed by vvwvv. h parc Please visit us on the web ! Scanned by on-line curator: Tony Gerbic * For FREE Distribution Only * RadioFans.CN 收音机爱 好者资料库 w $ 1 b : . .P, _ , 1, . ,. ,* 1. 721 A POWER SUPPLY OPERATING AND SERVICING MAN
2、UAL RadioFans.CN 收音机爱 好者资料库 0,PERATING AND SERVICING MANUAL FOR MODEL 721A POWER SUPPLY SERIAL 1 AND ABOVE Copyright HEWLETT-PACKARD COMPANY 1959 275 PAGE MILL ROAD, PAL0 ALTO, CALIFORNIA, U. S. A. 721A001 RadioFans.CN 收音机爱 好者资料库 Model 721A SPECIFICATIONS REGULATEDOUTPUT VOLTAGE: FULL LOAD OUTPUT CU
3、RRENT: LOAD REGULATION: RIPPLE AND NOISE: OUTPUT IMPEDANCE: METER RANGES: OVERLOAD PROTECTION: OUTPUT TERMINALS: POWER: WEIGHT DIMENSIONS: 0 to 30 volts dc, continuously variable. 150 ma. With the meter monitoring voltage, the changein output voltage from nc load to full load is less than 0 . 3 % or
4、 30 mv whichever is greater a 1 any output. Less than 150 pv rms. Less than 0.2 ohm in series with less than 30 ph, output terminal shunted by 0.1 pf. Full scale indications of: loma, 30 ma, 100 ma, 300 ma, 10 v and 30 v Maximum current selected by switch in four steps, 25 ma, 50 mal 100 ma and 225
5、ma. Three banana jacks spaced 3/4 in. apart. Positive and negative termin- als are isolated from chassis. A maximum of 400 volts may be con- nected between ground and either output terminal. 115/230 volts *lo%, 50-60 cps, 1 6 watts. Net 4 lbs., shipping 7 lbs. 7 in. wide, 4-3/8 in. high, 5-1/4 in. d
6、eep. Model 721A CONTENTS SECTION I GENERAL INFORMATION Page 1 . 1 Manual Content . i . 1 1 . 2 General Description . i . 1 1 - 3 Inspection . 1 - 1 1 - 4 Powercable . 1 - 1 1 . 5 230 Volt Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 2 SECTION I1 OPERATING INSTRUCTIONS 2 . 1 Oper
7、ating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . I1 . 1 2 . 2 Meter Range Switch . . . . . . . . . . . . . . . . . . . . . . . . . I1 . 1 Short Circuit Current Switch . Series Operation of Supplies . Parallel Operation of Supplies . 2 . 3 2 . 4 2 . 5 I1 . 1 I1 . 1 I1 . 3 SECTION I1
8、1 THEORY OF OPERATION 3 . 1 General Circuit Description . 111 . 1 Main and Auxiliary Supply Description . . . . . . . . . . . . . . . . 3 . 3 Reference Voltage . I11 . 1 3 . 4 Regulation Cycle Description . I11 . 1 3 . 5 Short Circuit Current Limiting Circuit . . . . . . . . . . . . . . . I11 . 1 Ou
9、tput Surge Protection Circuit . 3 . 7 Frequency Response Control . . . . . . . . . . . . . . . . . . . . . 111 . 2 3 . 2 I11 . 1 3 . 6 111 . 2 SECTION IV MAINTENANCE 4 - 1 4 - 2 4 - 3 4 - 4 4 - 5 4 - 6 4 - 7 4 - 8 4 - 9 4 . 10 Contents . IV . 1 General Maintenance Information . IV 01 Trouble Localiz
10、ation . IV . 1 Checking Voltage Regulation and Ripple . IV . 3 Measuring AC Internal Impedance . . . . . . . . . . . . . . . . . . IV . 3 Meter Calibration . IV . 3 Setting Maximum Output Voltage . IV . 5 Calibrating the Short Circuit Current Circuit . . . . . . . . . . . . IV . 5 Replacing the Powe
11、r Transistor . IV . 5 Replacing Diodes CR5, CR6 and CR7 . IV . 5 SECTION V TABLEOFREPLACEABLEPARTS 5 . 1 Table of Replaceable Parts . Model 721A Sect. I Page 1 1-1 MANUAL CONTENT The material for this instruction manual is written in five sections: Section I contains material of a general nature. Se
12、ction I1 explains how to operate the power supply. Section I11 explains how the circuit operates. Section IV covers maintenance and trouble shooting procedures. Section V is a table of replaceable parts. 1-2 GENERAL DESCRIPTION The Model 721A Power Supply produces a dc regulated voltage adjustable f
13、rom 0 to 30 volts. The supply makes load circuit performance inde- pendent of external power supply influences. The supply has very low source impedance and ex- cellent regulation against change in line and/or load. This supply is especially useful as a source of power for transistor circuits. A cir
14、cuit is provided which electronically limits the maximum output cur- rent that can be supplied to four nominal values selected by a front panel switch. This feature helps prevent the accidental destruction of an expensive transistor should an accident occur that would nor- mally allow excessive curr
15、ent to flow through it. The SHORT CIRCUIT CURRENT switch can be set to the value which is closest above the normal operating current. The supply will automatically limit the peak current flow to this nominal value regardless of the load resistance. Built-In Metering A built-in meter allows either ou
16、tput voltage or current to be monitored as selected by theMETER RANGE switch. SECTION I GENERAL INFORMATION Isolated Output The power supply has both output terminals in- sulated from chassis ground. Either terminal may be grounded or a number of supplies may be con- nected in series to obtain highe
17、r voltages. Insula- tion is such that the supply may be operated as high as 400 volts off of ground. Parallel ODeration Parallel operation of two or more supplies is pos- sible due to the unique electronic current limiting switch. The supplies will each contribute only the number of milliamperes sel
18、ected by the SHORT CIRCUIT CURRENT switch. The individual supplies may be loaded to approximately 225 ma with some reduction in ripple and regulation characteristics. Reliability The Model 721A Power Supply is very compact, and has low internal losses., which are made possible by fully transistorize
19、d circuitry. The trouble free characteristics of transistors together with the use of high quality components throughout, will result in a minimum of maintenance. 1-3 INSPECTION When the Model 721A is received, inspect it for damage received in transit. Operate the instrument to make certain that it
20、 is functioning satisfactorily. If damage is evident, follow the procedures outlined in the “CLAIM FOR DAMAGE IN SHIPMENT” page of this manual. 1-4 POWER CABLE The power cable consists of three conductors and is terminated in a three-prong male connector rec- ommended by the National Electrical Manu
21、fac- Sect. I Page 2 Model 721A turers Association. The third contact is an offset round pin added to a standard two-blade connector which grounds the instrument chassis when used with an appropriate receptacle. To use thisNEMA connector in a two-contact receptacle, a three- prong to two-prong adapte
22、r should be used. When the adapter is used, the third contact is terminated in a short lead from the adapter which can be con- nected to the outlet mounting box in order to ground the instrument cabinet. 1-5 230 VOLT OPERATION This instrument may be easily converted from 115 to 230 volt operation by
23、 removing two jumpers and installing one jumper. This changes the dual 115 volt primary windings from a parallel to a series connection. Refer to the schematic diagram and Figure 4-6 for details. The main fuse should be changed from 1/4 ampere slow-blow type to 1/8 ampere slow-blow type. I Model 721
- 配套讲稿: