1、 41 2A-6 - I MODEL 412A DC VOLTMETER-OHMMETER-AMMETER RANGE SWITCH MAINTENANCE The following precaution should be exercised when or organic solvents such as tolulene, carbon cleaning the 412A VTVM range switches. The Pyro- tetrochloride, or chlorinated compounds may cause film resistors have a silic
2、on coating - be careful to the wafer material to swell, thus should not be used. not wash the silicon off these resistors when cleaning We recommend use of either Freon TF or a soap and warm water wash. The wash must be followed by a the range switch. If you should accidentally remove the silicon co
3、ating it may be reapplied by using complete rinsing, followed by a rinse with distilled Dupont Silicon Insulcone. wat%r and overnight drying in an oven at approximately 120 F. - 412A-19W and 412A-19W-1 (White Wafers) This switch should be cleaned with distilled water or with Freon TF and completely
4、dried off. Any other cleaning agents may contain harmful chemicals which could attack the wafer material causing it to crack or lose its high impedance characteristics. w You may notice some cracks in the range switch. These should be cleaned; however, the cracks themselves are not detrimental to th
5、e switch operation unless the wafer has been broken. When using the soap and water cleaning process, it is imperative that all residue be rinsed free of the wafers to eliminate leakage paths between wafer contacts which will show up as a defect in the instru- ment operation. Extreme caution should b