1、I . : HP Archive . . This vintage Hewlett Packard document was preserved and distributed by www. Please visit us on the web ! . . On-line curator: Tony Gerbic RadioFans.CN 收音机爱 好者资料库 412A DC VACUUM TUBE VOLTMETER SERIALS PREFIXED: 004- RadioFans.CN 收音机爱 好者资料库 OPERATING AND SERVICING MANUAL MODEL 412
2、A DC VACUUM TUBE VOLTMETER SERIALS PREFIXED: 004- Copyright HEWLETT-PACKARD COMPANY 1959 275 PAGE MILL ROAD, PALC ALTO, CALIFORNIA, U. S. A. RadioFans.CN 收音机爱 好者资料库 Page 1 of 2 SPECIFICATIONS Model 412A TOLTMETER Voltage Range: Accuracy: Input Resistance: AC Rejection: MMETER Current Range: Accuracy
3、: Input Resistance: Positive and negative voltages from 1 millivolt full scale to 1000 volts ful scale in thirteen ranges. *l% of full scale on any range. 10 megohms *l% on 30 megohms *l% on 30 mv range. 100 megohms +l% on 100 mv range. 200 megohms *l% on 300 mv range and above. 1 mv, 3 mv, and 10 m
4、v ranges. A voltage at power line or twice power line frequency 40 db greater than ful scale affects reading less than 1%. Peak voltage must not exceed 1500 volts Positive and negative currents from 1 microampere full scale to 1 amperc full scale in thirteen ranges. *2% of full scale on any range. I
5、nternal Shunt Full Scale Internal Shunt Full Scale Range Resistance* Voltage Drop Range Resistance* Voltage Drop .001 ma 1000 ohms 1 mv 1 ma 1 ohm 1 mv .003 ma 316 ohms 0.9486 mv 3 ma 0.316 ohm 0.9486 mv .01 ma 100 ohms 1 mv 10 ma 0.1 ohm 1 mv .03 ma 31.6 ohms 0.9486 mv 30 ma 0.1 ohm 3 mv .1 ma 10 o
6、hms 1 mv 100 ma 0.1 ohm 10 mv .3 ma 3.16 ohms 0.9486 mv 300 ma 0.1 ohm 30 mv 1000 ma 0.1 ohm 100 mv * For total insertion resistance add 0.07 ohms copper lead resistance at 25OC. 7 Model 412A SPECIFICATIONS (CONTD.) Page 2 of 2 OHMMETER Resistance Range: Resistance from 1 ohm centerscale to 100 mego
7、hms centerscale in nine decade ranges. *5% of reading from 0.2 ohm to 500 megohms. 510% of reading from O.lto 0.2 ohm and from 500 megohms to 5000 megohms. Accuracy: Open Circuit Short Circuit Open Circuit Short Circuit Voltages and Currents: Range Volts Current Range Volts Current x1 10 mv 10 ma Xl
8、OK l v 100 Fa x10 100 mv 10 ma XlOOK l v 10 Fa XlOOO l v 1 ma XlOM l v .1 pa XlOOM l v .01 pa XlOO l v 10 ma X1M l v 1 Fa AMPLIFIER Voltage Gain: AC Rejection: output: Output Impedance: Noise: Drift: GENERAL Isolated Input: Power: Dimens ions : 1000 maximum Approximately 3 db at 1 cps, 80 db at 50 a
9、nd 60 cps. Proportional to meter indication; 1 volt at full scale. (Full scale corre- sponds to 1.0 on upper scale.) Less than 2 ohms at 0 cps. Less than 0.1% (rms) of full scale on any range. Negligible. Input terminals are isolated from case (power line) ground by a minimum leakage resistance of 1
10、00 megohms shunted by 0.1 Ff. Maximum potential difference between common Iead and case ground: 500 v peak. 115/230 volts *lo%, 50-60 cps, 35 watts. Cabinet Mount: 11-1/2 in. high, 7-1/2 in. wide, 10 in. deep. Rack Mount: 5-1/4 in. high, 19 in. wide, 10 in. deep. Weight: Cabinet Mount: Net 12 Ibs.,
11、shipping 17 Ibs. Model 412A CONTENTS SECTION I GENERAL DESCRIPTION General . 1 - 1 1 - 3 Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2 Damage in Shipment . 1 - 4 230 Volt Operation . . . . . . . . . . . . . . . . . . . . . . SECTION I1 OPERATING INSTRUCTIONS 2 - 1 Low-Level Ele
12、ctrical Phenomena . . . . . . . . . . . . . 2 - 2 Operating Instructions . SECTION I11 CIRCUIT OPERATION 3 - 1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2 Circuit Operation . . . . . . . . . . . . . . . . . . . . . . SECTION IV MAINTENANCE 4 - 1 4 - 2 4 - 3 4 - 4 4 - 5 4 - 6
13、4 - 7 4 - 8 4 - 9 4 . 10 4 . 11 4 . 12 Introduction . Required Test Equipment . . . . . . . . . . . . . . . . . . Performance Check . Cabinet Removal . Adjustments . Tube Replacement . Trouble Shooting . Modulator Check . Modulator Replacement . . . . . . . . . . . . . . . . . . . Demodulator Check
14、. Demodulator Replacement . . . . . . . . . . . . . . . . . Cable Replacement . SECTION V TABLEOFREPLACEABLEPARTS 5 - 1 Table of Replaceable Parts . . . . . . . . . . . . . . . . . Page 1 - 1 1 - 1 1 - 1 1 - 1 I1 . 1 I1 . 1 I11 . 1 I11 . 1 IV . 1 IV . 1 IV . 1 IV . 3 IV . 3 IV . 4 IV . 4 IV . 4 IV .
15、 5 IV . 5 IV . 6 IV . 6 v - 1 Model 412A 1-1 GENERAL The Model 412A DC . xuum Tube Voltmeter is a precision, wide range, multipurpose instrument which covers the entire range of voltage, cufrent, and resistance measurements normally encoun- tered in electronic equipment. It measures volt- ages from
16、less than 0.1 millivolt to 1000 volts, currents from less than 0.1 microampere to lam- pere, and resistances from 0.02 ohm to 5000 meg- ohms. 1-2 DAMAGE IN SHIPMENT Inspect and operate this instrument upon receipt. Section IV includes a performance check which is a good test as part of incoming qual
17、ity control in- spection. The check can be made with the instru- ment in its cabinet. If there is any damage, see the “Claim for Damage in Shipment” paragraph at the rear of this manual. 1-3 POWER CABLE The three-conductor power cable supplied with this instrument terminates in a polarized three-pro
18、ng Sect.1 Page 1 SECTION I GENERAL DESCRIPTION male connector recommended by the National Elec- trical Manufacturers Association. The third con- tact is an offset round pin added toa standard two- blade connector. This contact grounds the instru- ment when used with an appropriate receptacle. An ada
19、pter should be used to connect the NEMA plug to a standard two-contact output. When the adapter is used, the ground connection becomes a short lead from the adapter. This lead should be connected to a suitable ground for the protection of operating personnel. 1-4 230 VOLT OPERATION The 412A is norma
20、lly wiredat thefactoryfor oper- ation from a 115 volt, 50-60 cps power source. To convert it for use from a 230 volt, 50-60 cps power source, change the dual 115 volt primary windings of the power transformer from a parallel combination to a series combination. See sche- matic diagram for details. A
21、t the time of conver- sion, change line fuse from 1/2 ampere, slow- blow type to 1/4 ampere, slow-blow type. Model 412A Sect.11 Page 1 SECTION II 2-1 LOW-LEVEL ELECTRICAL PHENOMENA Stray low-level electrical phenomena are present, in one form or another, in nearlyall electrical cir- cuits. The 412A
22、does not distinguish between stray and signal voltages; it measures net voltage. Thus, when using the lower voltage ranges, consider the possibility of low-level electrical phenomena. Thermocouples (thermoelectric effect), flexing of coaxial cables (triboelectric effect), apparent re- sidual charges
23、 on capacitors (dielectric absorp- tion), battery action of two terminals mounted on an imperfect insulator (galvanic action) all can pro- duce voltages within the range of the 412A. The 412A voltage probe, current/resistance lead, and common lead are designed to have a very low thermoelectric effec
24、t with copper, the most com- mon electrical conductor. However, you may en- counter other materials. For example, the leads of many transistors are made of a mixtureof iron, nickel, and cobalt known commercially as Kovar, Fernico, etc. This material makes a very good thermocouple with copper: about