1、412A-5 MODEL 412A/AR VOLTMETER-OHMMETER-AMMETER SERIALS 134-07482 AND BELOW INPUT CABLE MODIFICATION KIT J STOCK NO. 412A-95B Installation of this modification kit will prevent damage to the input cables by reducing the effects of excess strain. A s a result, you will experience longer and more reli
2、able operation of the Model 412A/AR. Remove and discard bracket, This Service Note is written in two sections. The first section concerns kit installation instructions for cabinet model instruments. The second section It may be necessary to carefully separate concerns rack mount instruments. two res
3、istors to remove the inside nut. 4. Using a 1/4,) drive ratchet wrench, 5/16) socket, and a short extension, remove the two 5/16” nuts which hold the tie point bracket to the main chassis. Note Special tools other than an electric drill and a 1/4” 5. Remove original input cables and boots from drive
4、 ratchet wrench with a 5/16) socket are not front panel. Do not discard the voltage cable required. assembly . PARTS SUPPLIED IN KIT 412A-95B 6. Install three new boots on the front panel. Quantity Description Stock No. 7. Install new tie point bracket using the original #6 bracket screws (Refer to
5、Figure 1). 8. indicated in Figure 2. 1 Cable Assembly-Volt . . . . 412A-21D-1 Solder wires (not input cables) to tie points as 1 Cable Assembly-Common . . 412A-21E 1 Cable Assembly-Ohm-ma . . 412A-21F 6 1 Bracket, tie point . . . . . . 412A-12E of cable through boot. 1 Bracket, cable support, 1 Brac
6、ket, cable support, 3 Boot, cable. . . . . . . . . . 412A-83A nosed pliers may be used. 4 2 Hex Nut #6 . . . . . . . . . . 2420-0001 9. Install new ohms input cable by passing it through * * * * 428A-21A-1 the appropriate boot into instrument. Pull 12 inches (cabinet model) . . . . . . . 412A-12C 10
7、. Place the cable-clamp halves together on ohms cable 2 inches back from stripped end. Pass cable through the appropriate hole in cable support bracket (rackmount model). . . . . 412A-12D and press clamps firmly into hole. Heavy needle- Machine screw, 6-32 x 3/8 BH, withlockwasher . . . . . . 2390-0
8、009 11. Repeat steps 9 and 10 for new “volts” - and “common” cables. 12. Pull the cables forward through the boots. SECTION I, INSTALLATION INSTRUCTIONS CABINET MODEL 1. Remove cabinet and turn instrument on its top. 2. Remove shield on the function switch side of the instrument by removing the thre