HP-1728A-Manual-Display-Scanner 电路图.pdf
《HP-1728A-Manual-Display-Scanner 电路图.pdf》由会员分享,可在线阅读,更多相关《HP-1728A-Manual-Display-Scanner 电路图.pdf(48页珍藏版)》请在收音机爱好者资料库上搜索。
1、1-F.WLE:T -PACKARD COMPANf 1782A DISPLAY SCANNER RadioFans.CN 收音机爱 好者资料库 01550 1 OPERATING AND SERVICE MANUAL MODEL1782A SERIALS PREFIXED: 305- DISPLAY SCANNER Copyright HlWllTT-PACKAIO COMPANY 1963 1$01 PAOl Mill IOAO, PALO ALTO, CAllfOINIA, U.S.A. Printed: FEB 1963 RadioFans.CN 收音机爱 好者资料库 it Some
2、Model 175A Oscilloscopes with serial prefixes below 305-require modification in order to operate the Model 1782A Display Scanner. Please contact your nearest Hewlett-Packard representative for information regarding the changes. Model 1782A 01550-1 ; ; RadioFans.CN 收音机爱 好者资料库 . Model 1782A Table of C
3、ontents List of Illustrations and Tables TABLE OF CONTENTS Section Title Page I GENERAL INFORMATION . 1-1 1-1. Description . . . . . 1-1 1-3. Pen Stabilizer . 1-1 1-5. Marker . 1-1 1-7. Digital Recording . . . . . . . . . 1-1 1-9. Serial-Number System . . 1-1 Il PREPARATION FOR USE . 2-1 2-1. Incomi
4、ng Inspection . 2-1 2-2. Mechanical Inspection . 2-1 2-4. Performance Check 2-1 2-6. Power Requirement . 2-1 2-8. Repackaging for Shipment 2-1 Ill OPERATION . 3-1 3-1. Controls and Terminals . 3-1 3-3. Function SWitch . 3-1 3-8. Scanning SWitch . 3-1 3-10. Recorder Characteristics . 3-2 3-11. X-Y Re
5、corders . 3-2 3-13. Strip-Chart Recorders . 3-2 3-15. Operating Instructions . 3-2 3-19. Preliminary Adjustments 3-2 3-21. Step-By-Step Procedures 3-2 IV PRINCIPLES OF OPERATION . 4-1 4-1. Over-All Functional Description . 4-1 4-6. Circuit Description . . . . . . . 4-1 4-7. Integrator . . . . . . .
6、. . . . . . 4-1 4-12. Comparator . 4-2 4-14. Marker Output Circuit . . . . . . 4-2 Section Title Page IV PRINCIPLES OF OPERATION (Contd) 4-16. Sampler Circuit 4-2 4-18. Stretcher Circuit 4-3 4-21. Stabilizer Circuit 4-3 V MAINTENANCE 5-1 5-1. Performance Check 5-1 5-3. Test Equipment . 5-1 5-5. Prel
7、iminary Settings 5-2 5-7. Horizontal Output 5 -2 5-11. Vertical Output . . . . . . 5-2 5-14. Adjustments 5-3 5-16. Test Equipment 5-3 5-18. Preliminary Settings . . 5-3 5-19. Stretcher Gate Width 5-3 5-20. Comparator Compensation 5-3 5-21. Scan Start and Termination 5-4 5-22. Horizontal DC Level . .
8、 . . . 5-4 5-23. Input Attenuator Compensation . 5-4 5-24. Sampling Efficiency and Vertical Feedback . . . . 5-4 5-25. Sampling Gate Neutralization 5-5 5-26. Stabilizer DC Balance 5-5 5-27. Repairs . 5-5 5-29. Equipment Required 5-5 5-32. Procedure . . 5-5 5-34. Troubleshooting . . . . . . 5-6 5-35.
9、 Isolating Troubles to a Major Section 5-6 5-3 7. Circuit Troubleshooting 5-6 VI REPLACEABLE PARTS 6-1 6-1. Introduction . . . . . . 6-1 6-4. Ordering Information . . . . . 6-1 LIST OF ILLUSTRATIONS Number Title Page 1-1. Model 1782A Display Scanner . 1-0 3-1. Model 1782A Controls and Terminals . 3-
10、3 3-2. Recorder Calibration . . . . . . . 3-4 3-3. Automatic Scanning . . . . . . . . . . 3-5 3-4. Manual Scanning . 3-6 3-5. External Scanning . 3-7 4-1. Model 1782A Overall Block Diagram . 4-0 4-2. Model 1782A Simplified Block Diagram 4-1 4-3. Miller Integrator Simplified Circuit 4-2 5-1. Special
11、Attenuator . 5-1 5-2. Al Sampler Assembly, Location of Adjustments and Test Points . 5-3 Number Title 5-3. A2 Scan Generator Assembly, Location 5-4. 5-5. 5-6. 5-7. 5-8. 5-9. of Adjustments and Test Points . Vertical Feedback Adjustment . Sampler Storage Capacitor Adjustment Component Locations, Al S
12、ampler Assembly . Sampler Waveforms . Sampler Schematic . Component Locations, A2 Scan Generator Assembly . 5-10. Component Locations, Bottom View 5-11. Scan Generator Waveforms . 5-12: Scan Generator . . LI ST OF TA B L E S Number Title Page 1-1. Specifications . 1-1 5-1. Test Equipment for Check a
13、nd Adjustment . . . . . . . 5-0 5-2. Circuit Troubleshooting 5-6 6-1. Reference Designation Index 6-2 6-2. Replaceable Parts 6-7 Page 5-4 5-4 5-5 5-8 5-8 5-9 5-10 5-10 5-10 5-11 01550-1 iii Section I Figure 1-1 1-0 UNCT ON ARM RECORDE!lRECORD OFF SCAN START LOCK .a ,_ . MANUAL SCAN :-i . - DISPLAY l
14、iai SCANNER MODEL /!f) 178ZA VERT tiORIZ. AMPLITUDE l I .; . I OUTPUT l 0 SCANNING STAB LIZED LINEAR MANUAL. EXTERNAL : Figure 1-1. Model 1782A Display Scanner Model 1782A 01550-1 Model 1782A Section I Paragraphs 1-1to1-10 SECTION I GENERAL INFORMATION 1-1. DESCRIPTION. 1-2. The Model 1782A Display
15、Scanner provides outputs necessary to drive X-Y recorders so that the waveform displayed on the screen of Model 175A Oscil- loscope may be reproduced. Automatic scan and pen control are provided, with manual reset. Horizontal and vertical amplitude may be adjusted independently of zero. The effectiv
16、e bandwidth of the scanner is 30 me. Through the use of sampling techniques, the Model 1782A transforms high-speed phenomena to the bandwidth of conventional X-Y recorders such as the F. L. Moseley 3S or 2D. Sampling density is high enough to give a continuous presentation on the X-Y recorder. 1-3.
17、PEN STABILIZER. 1-4. An electronic pen stabilizer circuit provides a nearly constant writing rate to the recorder so that fast pulses may be recorded faithfully in a minimum of time. The pen stabilizer slows the horizontal pen speed during vertical deflection so that fast rise and fall time may be r
18、ecorded in detail. When there is no vertical deflection, the stabilizer increases horizontal pen speed to reduce scanning time. 1- 5 . MARKER. 1-6. With each sweep of the trace on the screen of the oscilloscope, the Model 1782A samples the am- plitude of the waveform at a single point and intensifie
19、s the trace beyond that point. Therefore the start of the intensified trace forms a marker which indicates the point being sampled at any particular instant. 1-7. DIGITAL RECORDING. 1-8. The Model 1782A may be used with auxiliary equipment to digitize the CRT display. By applying the appropriate ext
20、ernal scanning signal, the desired portion of the trace can be scanned either once or repeatedly, and the output reduced to digital form for recording or analysis. For example, an entire pulse train or a single pulse can be continuously monitored and the results permanently recorded in digital form.
21、 1-9 . SERIAL-NUMBER SYSTEM. 1-10. Hewlett-Packard instruments carry a two- section eight-digit serial number (000-00000). The first three digits are the serial prefix. If the serial prefix on your instrument does not agree with that shown on the title page of this manual, change sheets included wit
22、h this manual will describe the difference between the manual and your instrument. If the serial prefix on your instrument is not mentioned in the change sheets, obtain correct information through your Hewlett-Packard representative. Table 1-1. Specifications VERTICAL OUTPUT: HORIZONTAL OUTPUT: 0155
23、0-1 BANDWIDTH: SCANNING: SCANNING TIME: OSCILLOSCOPE SWEEP SPEED: REMOTE PEN LIFT: POWER: WEIGHT: 200 mv/ cm. Amplitude and de level are independently adjustable. Level: Adjustable to 0 v Amplitude: Adjustable from 0 to +15 v Greater than 30 me when installed in the Model 175A Oscilloscope with a 40