1、 MODELS 400D, 400H, AND 400L VACUUM TUBE VOLTMETERS TEST, ADJUSTMENT, AND CERTIFICATTON METHODS These Service Notes have been prepared as an aid in testing and adjusting all Model 400D, 400H, and 400L Vacuum Tube Voltmeters. The procedures that follow supersede those in other Service Notes or Op- er
2、ating and Servicing Manuals for these instruments. All circuit references apply to one of the three ref- erence schematic diagrams in these Service Notes. No attempt should be made to make an instrument agree completely with any one of these schematics. All recommended circuit changes are described
3、in Section I of these Service Notes. Voltages mentioned herein are measured between the indicated points and the chassis unless specified otherwise. The voltmeter power supply should al- ways be tested before undertaking repair and adjust- ment of any other section of the instrument. These Service N
4、otes are divided into five sections as follows: Section Page I MODIFICATIONS 1 I1 TUBE *CW* U Y W O R T *-I000 r(Mc.0 *db.,A. E* RadioFans.CN 收音机爱 好者资料库 Page 2 v- 400D-2C D. MODEL 400D SERIAL NO. 5130 AND BELOW Transpose the connections to electrolytic capa- citor C30 so that C30B is 20 pf and C3OC
5、is 40 pf. This transposition will reduce power supply ripple. E. MODEL 400D SERIAL NO. 14770 AND BELOW MODEL 400H S ERIAL NO. 1137 AND BELOW Replacement of control R29 or three-section resistor R28 in the amplifier feedback circuit will necessitate revising and rewiring the feedback circuit. The par
6、ts for this circuit revision are supplied as a kit with complete instructions under Stock No. 400D-95A. The revised feedback circuit is recommended for all 400D and 400H instruments whenever repair of the feedback circuit is necessary. The revised circuit is superior to earlier circuitry but does no
7、t offer suf- ficient advantages to warrant modification merely to include the new circuit in an instrument. The problem of adjusting frequency response following any changes in the feedback circuit would also make it advisable to avoid making unnecessary changes, F. MODEL 400D SERIAL NO. 16371 AND A
8、BOVE The regulated power supply was changed. Series regulator tube V7 was changed from a 6AU5 toa 12B4A. Control tube V8 was changed from a 6CB6 to a 6U8 and reference tube V9 was changed from an OB2 to a type 5651. The circuit differences can be determined by referring to the schematic diagrams inc
9、luded in these Service Notes. Field modification to incorporate this change is im- practical and is not recommended. SECTION II - TUBE stop when pointer is right on zero, If pointer overshoots zero, repeat this adjustment. 3) When pointer is exactly on zero, rotateadjustment screw approximately 15 d
10、egrees counterclockwise. This is enough to free adjustment screw from the meter suspension, If pointer moves during this step repeat entire adjustment. B. SETTING MECHANICAL ZERO IN MODEL 400L The special meter in the 400L is adjusted with the vtvm turned on since the meter pointer rests against the
11、 stop at the left end of the scale when the instrument is turned off. RadioFans.CN 收音机爱 好者资料库 Page 4 I In most 400L instruments the mechanical zero is pre- set during manufacture of the meter movement, The adjustment screw in such meters is inoperative. Most 400L instruments will have no adjusting s
12、crew on the front of the meter case. The following procedure is for use only with those instruments in which the mechanical adjustment screw is effective. 1) Connect a 1.0 volt rms, 400 cps signal from the Voltage Calibration Generator to the 400L INPUT terminals. Set the 400L to the 1-volt range an
13、d adjust the feedback control (R29) for a full scale meter reading of 1.0 volt. Any convenient range could also be used along with a corresponding change in input signal level. The input signal frequency is also not critical. 2) Switch the Calibration Generator todeliver a signal of 0.3 volt, or 0.3
14、 of the signal level used in step 1. Set the meter mechanical zero adjusting screw to ob- tain a reading that is 0.3 of the reference reading in step 1. Use the same vtvm rangefor both steps 1 and 2. If you overshoot, continue to rotate the screw clockwise and again approach the 0.3 point from the h
15、igh side of the scale, The screw should not be turned counterclockwise during this procedure. 3) Repeat steps 1 and 2 until no further improvement can be noted. Compromise adjustment of themechani- cal zero can be made for optimum tracking. After final adjustment, cover the screw with a small piece
16、of black tape. 1 I C. CHECK REGULATED POWER SUPPLY 1) The B+ voltage at the output of the regulated power supply should be 250 volts i5 volts. Measure this voltage between the chassis and the cathode of the series regulator tube with the line voltage set to 115 volts. 2) Vary the line voltage betwee
17、n 103 and 127 volts. The regulated B+ will usually change no more than 2 volts from one extreme to the other, Ripple volt- age in the regulated B+ is usually 3millivolts or less under these same test conditions. The dc voltage at the plate of the series regulator tube or at the cathode of the rectif
18、ier tube should be between 400 and 420 volts with a 115-volt line, Low voltage at this point may be from adefective rectifier tube, filter capacitor, or power transformer. To check the rectifier tube, connect a dc voltmeter be- tween chassis and the rectifier cathode. Reduce line voltage from 115 to
19、 103 volts. The voltmeter reading should drop as line voltage is reduced and then remain steady. If the dc voltage continues to drop at a slow rate, the rectifier tube is probably weak and should be replaced. If the regulated B+ voltage is not correct, replace- ment of any or all four tubes in the p
20、ower supply may be necessary. If tube replacement does not correct the dc voltage, precision resistors R62 and/or R64 may have changed value and should be replaced. Ab- normal current drain caused by component failure elsewhere in the vtvm can result in poor power sup- k ply regulation, 400D-2C Exce
21、ssive ac ripple may be from a defective tube in the power supply other than the rectifier tube, Ca- pacitor C36 may be open, D. SET DC HEATER VOLTAGE 1) Adjust control R66 to set the dc heater voltage for tubes V1 and V2 to 6.3 volts rt0.2 volt. The dc volt- ages across the heaters of tubes V3 and V
22、4 must also be 6.3 volts rt0.2 volt. DO NOT USE A VOLTMETER THAT IS CONNECTED TO A COMMON GROUND WITH THE VTVM UNDER TEST WHEN MEASURING THESE HEATER VOLTAGES. A SHORT CIRCUIT WILL RESULT. Slight readjustment of control R66 will often bring these two voltages within tolerance. 2) If both voltages ca
23、nnot be set to 6.3 volts, f 0.2 volt, adjust R66 to bring the lowest of the two voltages within tolerance. Connect a resistor in arallel with the two heaters with the highest voltage. )?he correct resistor value will enable you to bring bo* heater voltages within tolerance by readjusting control R66
24、. Use only one resistor in parallel with one pair of tubes at a time. A resistor value below 220 ohms 1 watt should not be used. 3) If you are unable to obtain the correct heater voltage after completing steps 1 and 2, checkelectro- lytic capacitors in the heater circuit. If both heater voltages are