1、MODEL 3484A MULTIFUNCTION UNIT RadioFans.CN 收音机爱 好者资料库 OPERATING AND SERVICE MANUAL MODEL 3484A MULTIFUNCTION UNIT -hp- Part No. 03484-90000 Serials Prefixed: 975- Copyright Hewlett-Packard Company 1970 P.O. Box 301, Loveland, Colorado, 80537 U.S.A. Printed: JULY 1970 RadioFans.CN 收音机爱 好者资料库 _J I N
2、0 0 I 0 N CJ 0STA0t! STAM! N!:Cf.SSAlV Ill MAILED 1 N UNITED STATES POSTAGE WILL BE PAID BY l . S T CLAS l - Model 3484A Section I SECTION GENERAL INFORMATION 1-1. INTRODUCTION. The hp Model 3484A Multifunction Unit, in conjunction with either the Model 3480A or 3480B Digital Voltmeter, makes 4-digi
3、t de voltage measurements, with up to 50% overrange capability. Full scale ranges of I 00 m V, IOOOmV, lOV, IOOV, and IOOOV may be selected manually, automatically, or remotely. Polarity selection and display arc automatic. The degree of filtering desired may also be selected manually or remotely. R
4、esistance and true rms ac voltage measurements may be made if Options 042 and 043 are added. 1-2. OPTIONS. 1-3. ISOLATED REMOTE OPTION 041. Option 041 provides remote programming connections which are isolated from the Model 3484A internal circuits. All external connections arc referenced to chassis
5、 (power line) ground. This option is available for field installation as -hp- Model 11151 A, Isolated Remote Assembly. Option 041 cannot be used unless the Model 3480A/B is equipped with Isolated BCD Output Option 004. 1-4. OHMS/DC CONVERTER OPTION 042. Option 042 allows the Model 3484A to make resi
6、stance measurements on six ranges of I 00 n, 1000 n, 10 kS1, 100kS1, 1000kS1, and JOMS1 full scale, with up to 50% overrange capability. Range may be selected manually, automatically, or remotely. This option is available for field installation as -hp- Model l l l 52A, Ohms/DC Converter Assembly. 1-
7、5. TRUE RMS AC CONVERTER OPTION 043. Option 043 allows the Model 3484A to make true rms ac voltage measurements on five ranges of 100 m V, 1000 m Y, 34114A MULTIFUNCTION UNIT FUNCTION RANGE SAMP RATE HOU).Jli T111a ZCRO JOY, l OOV, and IOOOV full scale, with up to 50% overrange capability. Range may
8、 be selected manually, automatically, or remotely. This option is available for field installation as -hp- Model 11 I 53A, True RMS AC Converter Assembly. 1-6. SPECIFICATIONS. Complete specifications for the Model 3484A and 3480A/B combination are given in Table 1-1. 1-7. INSTRUMENT AND MANUAL IDENT
9、IFICATION. Hewlett-Packard uses a two-section serial number. If the first section (serial prefix) of the serial number on your instrument does not agree with those on the title page of this manual, change sheets supplied with the manual will define the differences between your instrument and the Mod
10、el 3484A described in this manual. Some serial numbers may have a letter separating the two sections of the number. This letter indicates the country in which the instrument was manufactured. 1-8. ACCESSORY EQUIPMENT AVAILABLE. 1-9. The Model 1 I I 48A Plug-in Extender Cable allows the plug-in unit
11、to operate outside the 3480A/B to facilitate servicing. 110. The Model 11149A Remote Programming Cable is 6 feet Jong and is terminated at one end by a connector which mates with the 3484A rear panel remote connector Jl4. The other end of the cable is unterminated. 1-11. For additional information r
12、egarding optional equip- ment, contact your nearest -hp- Sales and Service Office, listed in Appendix B. FILTER TERMINAL Figure 1-1. Model 3484A Multifunction Unit. 1 -1 Section I 12 Model 3484A Table 1-1. Specifications. The following specifications apply to the Model 3484A Multifunction Unit used
13、in the Model 3480A/B Digital Voltmeter. DC VOLTS Full Scale Voltage Ranges 50% Overrange capability on all ranges; maximum input 1200 V. Range Selection Automatic Ranging Polarity Selection and Display Voltage Accuracy 24 hours (23C 1c, less than 50% relative humidity) 90 days (25C s0c, less than 95
14、% relative humidity) Temperature Coefficient o 0c to 55c Measuring Speed Reading Period Reading Rate (without range change) Automatic Manual Trigger External Trigger Response Time Autorange Time 100 mV, 1000 mV, 10 V, 100 V, 1000 V Manual, Automatic, or Remote Upranges at 140% of range Down ranges a
15、t 10% of range Automatic (0.01% of reading + 0.01% of range) 100 mV Range: (0.01 % of reading + 0.02% of range) All other ranges: (0.01% of reading+ 0.01% of range) 100 mV Range: Filter Out: (0.001% of reading + 0.0005% of range)/C Filter A or B: (0.001% of reading + 0.0015% of range)/C All other ra
16、nges: (0.001 % of reading + 0.0005% of range)/0c 950 microseconds Variable from 1 to 25 per second with front panel control Front panel pushbutton 0 to 1000 per second Without range change: Filter out: 1 msec to within 1 count of final reading when triggered coincident with step input voltage. Filte
17、r A : 200 msec to within 1 count of final reading Filter B: 1 second to within 1 count of final reading Filter out: 4 msec per range change Filter A : 200 msec per range change Filter B: 1 second per range change 1 1 Model 3484A Table 11. Specifications (Contd). DC VOL TS (Cont di Input Characterist
18、ics Input Resistance Effective Common Mode Rejection (Ratio of peak common-mode voltage to resultant error in reading with 1 kilohm unbalance in either lead) DC AC Normal Mode Rejection (Ratio of peak normal-mode signal to resultant error in reading) Filter Selection Zero Offset Voltage Stability (a
19、t constant temperature) Voltage Temperature Coefficient (0c to ss0c1 Current 12s0c s0c1 Current Temperature Coefficient 10c to ss0ci Noise Maximum Input Voltage Between HIGH and LOW terminals Between LOW and GUARD terminals Between GUARD and Chassis 100 mV, 1000 mV 8nd 10 V ranges: Greater than 101
20、ohms 100 V, 1000 V ranges: 10 megohms 0.1% Greater than 80 dB 50 to 60 Hz Filter out: Greater than 80 dB Filter A : Greater than 110 dB Filter B: Greater than 160 dB Filter out: 0 dB Filter A : Greater than 30 dB at 50 Hz and above Filter B: Greater than 80 dB at 50 Hz and above Manual or Remote Les
21、s than 10 microvolts per week Less than 1 microvolt per 0c Less than 10 pA Less than 1 pA per 0c Less than 20 microvolts peak-to-peak (unfiltered). Peak-to-peak noise is less than 20 microvolts 95% of the time since the noise amplitude approximates a Gaussian distribution where the standard deviatio
22、n (rms value) equals 5 microvolts. 1200 V peak 200 V peak 500 V peak Section I 1-3 Section I Model 3484A Table 11. Specifications. (Contd) 14 OHMS (OPTION 042) Full Scale Ranges 50% Overrange capability on all ranges Range Selection Automatic Ranging Measurement Accuracy 24 hours (23C 1c. less than
23、50% relative humidity) 90 days 12s0c s0c, less than 95% relative humidity) Measuring Speed Reading Period Reading Rate (Without range change) Automatic Manual Trigger External Trigger Response Time (Full scale step input, without range change) Autorange Time Input Characteristics Voltage across unkn
24、own resistance Maximum current through unknown resistance 100 ohms, 1000 ohms, 10 kilohms, 100 kilohms, 1000 kilohms 10 megohms Manual, Automatic, or Remote Upranges at 140% of range Down ranges at 10% of range 1000 ohm thru 1000 kilohm ranges: (0.01% of reading + 0.01% of range) 100 ohm range: (0,.