BK PRECISION 253X Instruction 电路图.pdf
《BK PRECISION 253X Instruction 电路图.pdf》由会员分享,可在线阅读,更多相关《BK PRECISION 253X Instruction 电路图.pdf(70页珍藏版)》请在收音机爱好者资料库上搜索。
1、 Digital Storage Oscilloscopes Model 2530 25 MHz, 250 MSa/s Model 2532 40 MHz, 500 MSa/s INSTRUCTION MANUAL Models 2530 Normal mode is best for most Pulse Width trigger applications Coupling DC AC HF Reject LF Reject 20 Falling edge Rising edge Triggerlevel can be adjusted vertically Figure 2-7 2.4
2、Acquisition system When you acquire a signal, the oscilloscope digitizes the signal and displays the resulting waveform on the screen. The acquisition mode defines how the signal is digitized, and the time base setting affects the time span and level of detail in the acquisition. The sampling rate a
3、djusts automatically to a value suitable for each selected Time/Div scaling factor. Refer to Appendix C for more detailed information. Push the ACQUIRE button to access the Acquisition menu Acquire button Figure 2-8 Note: Press the SINGLE button when you want the oscilloscope to acquire a single wav
4、eform Trigger coupling affects only the signal passed to the trigger system. It does not affect the bandwidth or coupling of the signal displayed on the screen. Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the video waveform has positive-going horizontal sy
5、nc pulses, use the Inverted Polarity selection. 21 Table 2- 6 Acquire Menu: The 3 acquisition modes Sample, Peak Detect and Average: ? Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to reconstruct the waveform. This mode accurately represents the s
6、ignal most of the time. Advantage: You can use this mode to reduce random noise. Disadvantage: This mode does not capture rapid variations in the signal that may occur between samples. This can result in aliasing and may cause narrow pulses to be missed. In these cases, you should use the Peak Detec
7、t mode to acquire data. Figure 2-9 Sampling mode ? Peak Detect: Peak Detect finds the maximum and minimum values of a signal. Advantage: In this way, the oscilloscope can acquire and display narrow pulses, which may have otherwise been missed in Sample mode. Disadvantage: Noise will appear to be hig
8、her in this mode. Menu Set up Explanation Sampling Use to acquire and accurately display most waveforms Peak Detect Use to detect glitches and decrease the possibility of aliasing. Average Select this mode to reduce the displayed random noise. Averages 4 16 32 64 128 256 Specify number of averages M
9、ode Equ time Real time Select realtime or equivalent sampling 22 Figure 2-10 Peak Detect mode ? Average: The oscilloscope acquires several waveforms, averages them, and displays the resulting waveform. Advantage: You can use this mode to reduce random noise. Figure 2-11 Average mode ? Equivalent Tim
10、e Sampling: The equivalent time sampling mode can achieve up to 20 ps of horizontal resolution (equivalent to 50 GSa/s). This mode is good for observing repetitive waveforms. ? Real Time Sampling: The maximum sampling rate for the 2530 is 250 MSa/s and 500MSa/s for the 2532. ? RUN/STOP Button: Press
11、 the RUN/STOP button when you want the oscilloscope to acquire waveforms continuously (Run/Stop button is green). Press the button again to stop the acquisition.( Run/Stop button is red) 23 ? SINGLE Button:Push the SINGLE button when you want the oscilloscope to acquire a single waveform and then st
12、op. (Make sure the Run Stop button is green and the display annunciator is in the Ready state). After the oscilloscope detects a trigger, it completes the acquisition then stops. Each time you push the SINGLE button, the oscilloscope begins to acquire another waveform. ? Base time: The oscilloscope
13、digitizes waveforms by acquiring the value of an input signal at discrete points. The time base allows you to control how often the values are digitized. Appendix C details the corresponding sampling rate for each time base setting. To adjust the time base to a horizontal scale that suits your purpo
14、se, use the Time/DIV knob. ? Time Domain Aliasing:Aliasing occurs when the oscilloscope does not sample the signal fast enough to construct an accurate waveform record. When this happens, the oscilloscope displays a waveform with a frequency lower than the actual input waveform, or triggers and disp
15、lays an unstable waveform. Sampled points Apparent low-frequency waveform due to aliasing Actual high-frequency waveform Figure 2-12 2.5 Display System Push the display button to choose how waveforms are presented and to change the appearance of the entire display. Display System Figure 2-13 24 Tabl
16、e 2-7 Figure 2-14 Figure 2-15 Options Set up Explanation Type Vectors Dots Vectors fills the space between adjacent sample points in the display Display sample points Persist Off 1 sec 2 sec 5 sec Infinite Sets the length of time each sampling point remains visible. Contrast Increase the display con
17、trast Decrease the display contrast Format YT XY YT format displays the vertical voltage in relation to time (horizontal scale) XY format displays a dot each time a sample is acquired on channel 1 and channel 2 Screen Normal Inverted set to normal mode set to invert color display mode Grid Display g
18、rids and axes on the screen Turn off the grids Turn off the grids and axes 25 2.5.1 X-Y Format Use the XY format to analyze phase differences, such as those represented by Lissajous patterns. The format plots the voltage on channel 1 against the voltage on channel 2, where channel 1 is the horizonta
19、l axis and channel 2 is the vertical axis. The oscilloscope uses the untriggered sample acquisition mode and displays data as dots. In this mode, the sampling default sampling rate is 1 MSa/s and is adjustable from 5kSA/s 200 MSa/s. In this mode, the controls operate as followed: The Channel 1“Volts
20、/div”and vertical “POSITION” controls set the horizontal scale and position. The Channel 2“Volts/div”and vertical “POSITION” controls continue to set vertical scale and position. The following functions do not work in XY mode: Reference or math waveform Cursors Autoset(resets display format to YT) T
21、rigger Controls Horizontal Position Knob Vector Display Type Scan Display 2.6 Measurement system There are several ways to take measurements. You can use the graticule, the cursors, or make an automated measurement. Note:The oscilloscope can capture a waveform in normal YT mode at any sampling rate.
22、 You can view the same waveform in XY mode. To do so, stop the acquisition and change the display format to XY. Note: When the display Type is set to Vectors, the oscilloscope connects the sample points by using digital interpolation. Interpolation mode contain linearity interpolation and sin(x)/x i
23、nterpolation. Sin(x)/x interpolation is suitable for real time mode and is most effective at 250ns/div or faster time. The oscilloscope automatically switches to the most suitable display type based on the timebase setting. 26 2.6.1 Graticule Measurement This method allows you to make a quick, visua