1、PN 778134 February 1986 Rev. 8, 7/00 1985, 1995, 1996, 2000 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companies. Introduction The Model 80T-150U Universal Temperature Probe is a self- contained temperature-to voltage converter. The pr
2、obe is designed to provide a direct temperature reading when it is connected to any high impedance DMM that is capable of 1 mV resolution and at least a 300-count full scale readout capability. Output is l mV per degree (Celsius or Fahrenheit). Two switch-selected temperature output scalings are pro
3、vided: -50 to +150C or -58 to +302F. The probe will stand off 60V dc or 30V ac rms (42.4V peak). The unit is housed in two separate but attached assemblies: a temperature probe and a temperature-to-voltage converter. The probe contains the temperature-sensing element and is electrically connected to
4、 the temperature-to-voltage converter through a 46- inch shielded cable. A three-position switch on the converter acts as a power switch and is used for selecting Celsius or Fahrenheit scaling for the output. Two banana plugs with standard 0.75-inch spacing are provided for connecting the 80T-150U t
5、o the DMM. Operating power for the 80T-150U is derived from a standard 9V battery. Typically, an alkaline battery provides more than 1600 hours of continuous operation before replacement is necessary. An OFF switch is provided on the temperature-to-voltage converter to allow battery conservation whe
6、n the unit is not in use. In addition, the OFF position of the power switch allows the battery condition to be determined via the external DMM. Temperature is measured by exposing the probe tip directly to the material to be measured (non-corrosive liquid, gas, or solid). A direct temperature readin
7、g is displayed on the DMM. Specifications The 80T-150U will achieve rated accuracy when it is used with any 0.25% DMM that has an input impedance of 1 M. Electrical Measurement Range: (See Probe Limitations) -50 to +150 degrees Celsius -58 to +302 degrees Fahrenheit Accuracy: (Applies for one year a
8、fter purchase or calibration) AMBIENT CACCURACY +15 to +35C 1C from 0 to +100C, decreasing linearly to 3C at -50 and +150C 0 to 15C and 2C from 0 to +100C, +35 to +50Cdecreasing linearly to 4C at -50 and +150C AMBIENT FACCURACY +59 to +95F 1.8F from +32 to +212F, decreasing linearly to 5.4F at -58 a
9、nd 302F +32 to +59 F and 3.6F from +32 to +212F, +95 to +122 Fdecreasing linearly to 7.2 F at -58 and 302F Sensitivity (80T-150U output): 1 mV dc / C or F Voltage Standoff: 60V dc or 30V ac rms (42.4V peak) Settling Time: 5.5 seconds to settle within 2 for a 50 change RF Field: Exposure to an RF fie
10、ld interferes significantly with this modules capacity to accurately measure temperature. The interference is temporary and the module suffers no harm when exposed to an RF field of 3 V/m. Accurate temperature measurement capability is completely restored when the module is removed from the RF field
11、. Common sources of RF fields are handheld two-way radios and cellular telephones. If such a source is suspected of interfering with this module, either turn off the transmitter or increase the separation between the transmitter and the module. Environmental Ambient Operating Range for Temperature M
12、odule: 0 to +50C (+32 to +122F) Maximum Temperature Probe Body and Cable: +70C (160F) See Probe Limitations Minimum Temperature Probe Body and Cable: Minimum Operating Temperature for Cable: -12C (10F) Minimum Operating Temperature for Probe Body: -40C (-40F) See Probe Limitations Storage Temperatur
13、e for Unit: -40 to +70C (-40 to +160F) Humidity: 0% to 90% (0C to 35C) 0% to 70% (35C to 50C) (noncondensing) Altitude:Operating: 10,000 feet Storage: 50,000 feet Application Force: 15 pounds maximum (probe tip to measured surface) General Weight: 5.7 ounces, 161.5 grams Overall Length: 53.8 inches,
14、 1.36 meters Battery: Standard 9V battery (NEDA #1604,6F22,006P) Battery Life: 1600+ hours, typical (Alkaline Battery), 6.5V minimum Output Termination: Standard 0.75-inch spaced double banana plug Probe Material: Glass-filled valox Probe Size: 0.6 maximum diameter Tip Material: Aluminum Tip Size: 0
15、/07 to 0.08 diameter, 30% convexed Operating Notes The following paragraphs are intended to familiarize the operator with the 80T-150U. The operator should read these paragraphs before attempting to operate the probe. Probe Limitations The 80T-150U probe is constructed of a highly durable plastic an
16、d is suitable for measuring the temperature of liquids, gases, and solid surfaces up to 150C. When measuring temperature, observe the following precautions to prevent damage to the probe: 1. Do not expose the probe end (probe tip plus about 2 inches of the probe body) to temperatures exceeding +150C
17、 (302F). The remainder of the probe body should not be exposed to temperatures above +70C (160F). 80T-150U Universal Temperature Probe Instruction Sheet RadioFans.CN 收音机爱 好者资料库 2. Do not expose the probe end (probe tip plus about 2 inches of the probe body) to temperatures below -50C (-58F). The rem
18、ainder of the probe body should not be exposed to temperatures below -40C (-40F). 3. For liquid measurements, recommended applications range from water, lubricants, and fuels to most solvents. Liquids as shallow as 1/2 inch can be measured since the temperature sensor is in the probe tip. Warning To
19、 avoid electrical shock, do not use this instrument when voltages exceeding 60V dc or 30V ac rms (42.4V peak ) are present. The probe tip is electrically connected to the output terminals. Caution Long-term exposure of the probe to corrosive environments will result in pitting and deterioration of t
20、he aluminum probe tip. Error Sources When the probe tip is applied to a solid surface, it draws or sinks heat from the surface. Therefore, if the measured surface has a low mass (e.g., a transistor case), the indicated temperature may be lower than the actual temperature. Similarly, a steady-state e
21、rror or gradient exists between the measured surface and the sensing device in the probe tip. This is due to the flow of heat from the measurement surface to the probe body. The effect of the steady-state error increases as the differential between ambient and surface temperature increases. To deter
22、mine the actual surface temperature of a device, both the heat-sinking and steady-state errors must be considered. The correction curve given in Figure 1 approximates the effect of both error sources on TO-3, TO-5, and TO-18 transistor cases. RF signals applied to the 80T-150U probe tip can also cau
23、se errors in temperature measurement. Figure 2 defines the rf signal limits that can be tolerated without degrading measurement accuracy. F1.eps Figure 1. Initial Case Temperature Above Ambient vs Meter Reading Above Ambient F2.eps Figure 2. Maximum Signal RF Limits (Vrms) at Probe Tip Operation Use
24、 the following procedure to operate the 80T-150U probe: 1. Connect the banana plugs on the 80T-150U to the input terminals of a high impedance DMM. Observe polarity. 2. Select a dc voltage range that will provide at least 1 mV resolution (1 mV/degree) and a full scale readout that will encompass the